Reference Manual
748384-C
September 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Analyzer Variables
2-41
NGA2000 Reference
OZON_SW
Direction Output
Binding
None
Type Enumerated
Value range
OFF; ON
Units None
Description
The state of the ozonator pressure switch
Usage
There is a pressure switch in the CLD that detects the presence of ozonator air (or
oxygen). The reaction that detects NO cannot occur in the absence of this.
See also
OZONATOR
OZONATOR
Direction Bidirectional
Binding
None
Type Enumerated
Value range
POWERED; UNPOWERED
Units None
Description
CLD only; the state of the ozonator lamp
Usage
Shows the state of, and may be used to control the state of the ozonator lamp. This
may be switched off to conserve its life, and also to conserve the life of the plumbing
components around it in the absence of ozonator air or oxygen flow. If a technician is
working on the unit, he should switch the lamp off so as to avoid both breathing ozone
and burning his eyes with the ultra-violet light emitted by the lamp. Ozone is very
harmful to lung tissue.
See also
OZON_SW
P_WAS
Direction Output
Binding
None
Type
4-element floating point array
Value range
Depending on configuration..
Units hPa
Description
FID only; manufacturing check out time values:
Sample capillary pressure.
Burner air pressure.
Fuel supply pressure.
Purge gas pressure.
Usage
Historical values of PRESS_IS. These show how the instrument was working when it
was tested during the manufacturing process. These values are loaded automatically
by the manufacturing data storage function.
See also
PRESS_IS
P2VFLL
Direction Bidirectional
Binding
None
Type
Floating point
Value range
0 - 1000
Units cc/min
Description
FID only; bypass sample flow lower limit
Usage
Used to generate a warning alarm if the bypass sample flow drops below the value
set in this variable.
See also
FLOW_IS