Analyzer Variables
Rosemount Analytical Inc. A Division of Emerson Process Management
2-52
NGA2000 Reference
SPAN
Direction Bidirectional
Binding
None
Type
4-element floating point array
Value range
Each element typically around 0.0005, depending on range and analyzer.
Units None
Description
The span coefficient array, element 1 being the value of the span coefficient for range
1.
Usage
The multiplicative factor that converts the raw signal minus the zero offset to the
normalized value between 0 and 1 for use in the linearizing function. This value is set
by the span function in the analyzer, though it may be overwritten by the user, or
copied over from the SPAN_WAS or SPAN_GOOD values. It may be copied into
SPAN_WAS or SPAN_GOOD by functions in the analyzer menus.
See also
SPAN_GOOD
SPAN_WAS
ZERO
SPAN_COMP
Direction Bidirectional
Binding
None
Type
Floating point
Value range
Around 0
Units None
Description
NDIR and PMD only; determines the temperature compensation slope coefficient.
Usage
Used to set the temperature compensation in the NDIR and PMD analyzer. This is
experimentally determined by first compensating the zero offset with ZERO_COMP,
and then running span gases through the analyzer while varying the temperature. It is
essential that the analyzer be linearized correctly beforehand. If the user is not in a
position to do this experiment, this value should not be adjusted
See also
ZERO_COMP
SPAN_GOOD
Direction Bidirectional
Binding
None
Type
4-element floating point array
Value range
Each element typically around 0.0005, depending on range and analyzer.
Units None
Description
A known good value of the span factor SPAN.
Usage
The user can copy a known good set of span coefficients into this variable, and then if
desired copy them back into SPAN. It is a good way of insuring against accidental
invalid calibrations.
See also
SPAN
SPANWAS
SPAN_THEN
Direction Output
Binding
None
Type
4-element floating point array
Value range
Equal to the span gas value when the unit was linearized
Units ppm
Description
The value of the span gases when the unit was linearized.
Usage
Used for the temperature compensation scheme, this value is recorded by the analyzer
so as to allow it to compensate correctly.
See also
SPANGAS