Reference Manual
748384-C
September 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Analyzer Variables
2-25
NGA2000 Reference
FID_MSGE
Direction Output
Binding
None
Type Enumerated
Value range
PARAMETERS NORMAL; FUEL PRESSURE TOO LOW; FUEL PRESSURE TOO
HIGH; AIR PRESSURE TOO LOW; AIR PRESSURE TOO HIGH; SAMPLE
PRESSURE TOO LOW; SAMPLE PRESSURE TOO HIGH; PURGE PRESSURE
TOO LOW; PURGE PRESSURE TOO HIGH; FUEL SELECTION IN-VALID; PURGE
SYSTEM FAILURE; IGNITION FAILED; OVEN TEMPERATURE TOO LOW; FLAME
THERMISTOR FAILURE; SYSTEM PURGING... PLEASE WAIT.
Units None
Description
FID only; reports the above list of states.
Usage
Reports the present status of the above factors
See also
DIGDIAG
DIGDIAGA
FLAME_ON_TIM
Direction Output
Binding
None
Type String
Value range
31 characters
Units None
Description
FID only; the length of time the flame has been on
Usage
Reports this time.
See also
FLAME_STATUS
Direction Output
Binding
None
Type Enumerated
Value range
Ready; Enrich fuel; Light flame; Retry; FAILURE; Flame on
Units None
Description
FID only; reports the state of the flame ignition cycle.
Usage
Reports the present status of the ignition cycle.
See also
DIGDIAG
DIGDIAGA
FLOW_FACTOR
Direction Bidirectional
Binding
None
Type Floating
point
Value range
1; can vary from 0.25 to 4
Units None
Description
CLD only; the capillary flow factor
Usage
Used as a multiplier for the CAP_FLOWRATE variable so that the output
compensation for flow changes is correct
See also
CAP_FLOWRATE