Reference Manual
748384-C
September 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Analyzer Variables
2-53
NGA2000 Reference
SPANGAS
Direction Bidirectional
Binding
None
Type
4-element floating point array
Value range
Typically 80% of RNGHI
Units ppm
Description
Contains the value of span gases, element 1 containing the span gas for range 1 etc.
Usage
During a span function, the analyzer adjusts the value of SPAN for the selected range
such that the value of PVA becomes equal to the value of the appropriate element of
SPANGAS, to within 0.02% of the range value. This variable should be kept updated
with the current value of the span gas as shown on the bottle label. If the bottle is of
dubious quality, it may be desirable to compare its reading with that of a known good
standard, and then mark the bottle and this variable accordingly.
See also
SPAN
SPANWAS
SPANWAS
Direction Bidirectional
Binding
None
Type
4-element floating point array
Value range
Each element typically around 0.0005, depending on range and analyzer.
Units None
Description
The value of the span factor SPAN at manufacturing checkout time.
Usage
The check out personnel copy a known good set of span coefficients into this variable,
and then if desired the user can copy them back into SPAN.
See also
SPAN
SPAN_GOOD
ST1NAME
Direction
Analyzer input
Binding
None
Type
Single element enumeration
Value range
NORMAL, MAINTENANCE REQUEST, FAILURE, CAL IN PROGRESS, ZERO IN
PROGRESS, SPAN IN PROGRESS, ZERO FAILED, SPAN FAILED, RANGE
OVERFLOW. RANGE UNDERFLOW, FLOW TOO LOW, FLOW TOO HIGH, INVALID
SELECTION, RANGE 1, RANGE 2, RANGE 3, RANGE 4
Units None
Description
Each bit of the first and second element of STCONT shows the state of the
corresponding alarm named in this variable; These are standard alarms supported by
each analyzer. MAINTENANCE REQUEST means that GENERALSTATE is greater
than 0; FAILURE that it is equal to' FAILURE. RANGE OVERFLOW means that the
reading is greater than the CURENTRNGHI value, RANGE UNDERFLOW that it is
less than the RNGHI value of the range below the present one. INVALID SELECTION
is not shown on the display.
Usage
Used to allow the 1/0 Module to select which specific alarm to assign to which relay.
These values are shown on the 1/0 module alarm selection menu.
See also
ST2NAME
ST3NAME
ST4NAME
STCONT