9
MODEL RDO
WIRING
2.1 ELECTROSTATIC DISCHARGE (ESD)
RECOMMENDATIONS
•
Before making wiring connections or
touching circuit boards or other internal
components, discharge any static elec-
tricity from your body by touching a grounded metal
object.
•
When making wiring connections, make sure to
remain properly grounded by wearing an ESD wrist
strap or similar device.
2.2 CUSTOMER-SUPPLIED ELECTRICAL
EQUIPMENT
The user must supply the following:
•
18-12 AWG wire for electrical connections using
conduit or 100-240 VAC power supply
• 24-12 AWG cable for connecting relays and PLC
devices
• A 100-240 VAC or 9-36 VDC source with over cur-
rent/disconnect protection for hard-wired locations
using conduit.
• Approved suitable wiring for cord-connected plug
and socket locations.
SECTION 2.0
WIRING
DANGER
Only properly trained and qualified personnel
should install the RDO analyzer described in this
manual. This instrument should be installed for use
in nonhazardous locations only.
DANGER
2.3 ENSURING GOOD ELECTRICAL
CONNECTIONS
To ensure that all sensors and power sources function
properly, make sure that:
•
Each individual wire is stripped and tinned to ¼ in.
•
Each wire is tightly screwed into the terminal strip.
•
Each wire is touching the terminal strip. If the plas-
tic wire jacket is clamped into the terminal strip, con-
nections will not be made.
•
Clip or cap any unused wires.
DANGER
Make sure that power to the instrument is discon-
nected before making any wiring connections.
DANGER
Débrancher toute alimentation à l'appareil avant de
connecter les fils.
DANGER
Do not connect low voltage circuits (<50 V) to the
terminal connectors on the AC board!
DANGER
Ne pas connecter des circuits basse tension
(moins de 50 V) aux bornes de connexion sur la
carte électronique de secteur (courant alternatif)!
2.4 AC POWER AND HIGH VOLTAGE
RELAY CONNECTIONS
L'installation de l'appareil "RDO" décrite dans le
guide d'utilisateur doit impérativement être réalisée
par des personnes qualifiées. L'installation de cet
appareil est seulement prescrite pour des emplace-
ments sans risques.
CONNEXIONS AUX COURANT SECTEUR /
ALTERNATIF
CONNEXIONS AUX RELAIS