29
MODEL RDO
SPECIFICATIONS
SECTION 8.0
SPECIFICATIONS
8.1 RDO PROBE
Wetted Materials:
Delrin
1
, ABS, Viton
2
, titanium, polycar-
bonate/poly(methyl methacrylate) blend
Dimensions:
length 8.0 in (203 mm) diameter 1.9 in
(47 mm)
Rating:
IP-67 with cap off; IP-68 with cap installed
Process connection:
1-1/4 inch FNPT
Integral cable length:
32 ft (10 m)
Maximum cable length (quick disconnect cable only):
4000 ft (1219 m)
Pressure:
up to 314 psig (2060 kPa abs)
Temperature:
32 to 122°F (0 - 50°C)
Range:
0 to 20 ppm (mg/L) or 0 to 200% saturation
Accuracy:
±0.1 ppm between 0 and 8 ppm; ±0.2 ppm
between 8 and 20 ppm
Resolution:
0.01 ppm (mg/L)
Digital output:
Modbus/RS485
Response time:
30 sec to 90% of final value; 37 sec to
95% of final value (at 25°C)
Operating life of sensing cap:
1 year from first reading
Shelf life of sensing cap:
2 years from date of manu-
facture; for full operating life install the sensing cap
within one year of manufacture date.
Safety directive:
73/23/EEC
EU directives:
2004/108/EC for Electro-
magnetic Compatibility (EMC) and
72/23/EEC for Safety
Immunity:
EN 61000-6-2, Electromagnetic Compatibility
(EMC) part 6-2
Emissions:
Class A requirements of CISPR 11:2004
8.2 RDO ANALYZER
Enclosure:
Polycarbonate, rated NEMA 4X, IP67
Dimensions (W x H x D):
6.3 x 6.3 x 3.6 in (16 x 16 x
9.0 cm)
Display:
liquid crystal; character height 0.4 in (6 mm)
Mounting:
suitable for pipe or wall mounting
Conduit openings:
six; PG 13.5 (
1
⁄
2
inch); three gland
fittings and five plugs ship with the analyzer
Ambient temperature and humidity:
-4 to 158°F (-20 to
70°C); 95% RH (non-condensing)
Power:
100 to 240 VAC, approximately 0.15 A, 50-60 Hz
Analog Outputs:
two fully scalable 4-20 mA outputs,
each loop-powered (9-36 VDC)
Digital Output:
Modbus/RS485
Relays:
two low voltage relays, <50 VAC or VDC, maxi-
mum current 2 A (resistive or inductive) two high
voltage relays, 264 VAC max, maximum current 5 A
(resistive or inductive)
Barometric pressure range:
8.86 to 29.53 in Hg (300 to
1000 mbar)
Barometric pressure accuracy:
±0.09 in Hg (±3 mbar)
EU directive:
2004/108/EC for Electromagnetic
Compatibility (EMC)
Immunity:
EN61000-6-2, Electromagnetic Compatibility
(EMC) part 6-2
Emissions:
EN61000-6-4 Electromagnetic Compatibility;
includes IEC/EN 61000-3-2 and IEC/EN61000-3-3,
where applicable
Safety:
UL 61010-1 and CAN/CSA C22.2 #61010-1
1 Delrin is a registered trademark of DuPont DeNemours, LLC
2 Viton is a registered trademark of DuPont Dow Elastomers, LLC
Summary of Contents for Rosemount Analytical RDO
Page 2: ...2 this page left blank intentionally ...