Rosemount LinkPro Operating Instructions
November 2017
Reference Manual
00809-0100-4570, Rev DA
Rosemount LinkPro Operating Instructions
179
1. Verify with the cellular provider that the SIM card supports GPRS data.
2. Verify with the cellular provider the APN, user name, and password (if required) needed to access the
GPRS data.
3. Ensure that the Rosemount LinkPro is configured with the User Name and Password for GPRS Internet
Access if required (use the Modem Wizard to configure the Rosemount LinkPro).
4. Connect the Rosemount LinkPro directly to the computer and run the Modem Wizard, fill all required
fields and select
Analyze
. Ensure that the analysis passes all tests. If not, consult with the cellular
provider or Emerson Technical Support. Ensure that the Rosemount 5708 is not connected to the
Rosemount LinkPro while analyzing.
5. Check the power supply connected to the Rosemount LinkPro, it should be 24Vdc with a limitation
above 1.5A.
6. Ensure that every Rosemount 5708 has a different polling address and one device use to polling
address 00.
7. Check the RS-485 connection between the Rosemount 5708 and the Rosemount LinkPro.
8. Check for mutual grounding of the RS-485 between the power supply, Rosemount LinkPro, and
Rosemount 5708.
9. Check the existence of 120
Ω
resistors on both ends of the RS-485 bus.
10. Ensure that cable resistance is 60Ù +/-5% between the D+ and D- ports. Ensure that there is no
connection to the Rosemount 5708 and the communication is idle.
11. Check the LED’s behavior on the modem. The green LED should be solid when the SIM card is
registered with the network.
12. Try calling the Rosemount LinkPro with the SIM card to check that the Rosemount LinkPro has
acquired the network.
13. Verify that the SIM card phone number is correct.
14. Verify that the SIM card receives SMS messages. This can be tested by inserting the SIM card into a
mobile phone and checking the options of receiving and sending SMS messages.
15. Send the following reset SMS to the Rosemount LinkPro: HW MODEM RESET (this is a case sensitive
message).
16. Ensure that the TCP port is open in the router and forwarded to your computer.
17. Verify that your computer still has the same External IP address.
18. Re-check the External IP address of the network your computer is connected to (in websites like:
www.whatismyip.com
).
Summary of Contents for Rosemount 5708
Page 1: ...Reference Manual 00809 0100 4570 Rev DA November 2017 Rosemount 5708 3D Solids Scanner ...
Page 2: ...ii Section Title November 2017 Reference Manual 00809 0100 4570 Rev DA ...
Page 8: ...vi Reference Manual 00809 0100 4570 Rev DA Contents November 2017 Contents ...
Page 12: ...2 Reference Manual 00809 0100 4570 Rev DA Introduction November 2017 Introduction ...
Page 202: ...ii Reference Manual 00809 0100 4570 Rev DA Index November 2017 Index ...
Page 203: ...iii Reference Manual 00809 0100 4570 Rev DA November 2017 ...