4. Check the instrument side connector to determine which pin (#1
or #2) is designated as the positive (+) pin. Insert the positive (+)
field side contact into the correct pin location of the crimp
terminal. Follow the same procedure for the negative (-) wire and
the ground wire (if applicable). Consult the factory regarding 4-
wire configurations.
IMPORTANT
Ensure the positive (+) and negative (-) cable wires are installed on
the correct pins on the field side connector to match the polarity
configuration of the instrument side connector.
5. Firmly push the contacts into the crimp terminal until a small snap
is heard and the contact is locked in place. Gently pull each wire
to confirm it is locked into the crimp terminal.
Note: If the contacts should be installed incorrectly, use the
contact removal tool to safely remove them. Push the tool into
the front side until the crimp contact is released.
1
6
3
8
4
7
5
2