Rosemount 356
Reference Manual
00809-0100-4856 Rev AE
February 2020
A-4
Prepare Cable and
Attach Crimp Contacts
1. Remove approximately 70 mm of the outer sheath of the cable,
taking care not to damage the metal braiding.
2. Cut and remove the metal braiding so that there is a minimum of
15 mm remaining.
3. Cut and remove the inner sheath to expose approximately 55 mm
of the conductor wires.
4. Cut any unused conductor wires flush with inner sheath. Strip
8 mm of insulation from the remaining wires, taking care not to
cut or damage individual strands. Gently re-twist the stranded
conductors if necessary.
70 mm
15 mm
8 mm
55 mm