HART Communicator
Out of memory for hotkey
configuration. Delete
unnecessary items.
There is no more memory available to store additional hotkey items.
Unnecessary items should be deleted to make space available.
Overwrite existing
configuration memory
Requests permission to overwrite existing configuration either by a
device-to-memory transfer or by an offline configuration. User answers using
the softkeys.
Press OK...
Select the
OK
softkey. This message usually appears after an error message
from the application or as a result of HART communications.
Restore device value?
The edited value sent to a device was not properly implemented. Restoring the
device value returns the variable to its original value.
ROM checksum error ON
Checksum of transmitter software has detected a fault. Return to Rosemount
Nuclear Instruments, Inc. for repair.
Save data from device to
configuration memory
Prompts user to select
SAVE
softkey to initiate a device-to-memory transfer.
Saving data to configuration
memory.
Data is being transferred from a device to configuration memory.
Sending data to device.
Data is being transferred from configuration memory to a device.
Sensor board not initialized
ON
The sensor module electronics board is not initialized. Return to Rosemount
Nuclear Instruments, Inc. for repair.
There are write only
variables which have not
been edited. Please edit
them.
There are write-only variables which have not been set by the user. These
variables should be set or invalid values may be sent to the device.
There is unsent data. Send it
before shutting off?
Select
YES
to send unsent data and turn the HC off. Select
NO
to turn the HC off
and lose the unsent data.
Too few data bytes received
Command returns fewer data bytes than expected as determined by the device
description.
Transmitter Fault
Device returns a command response indicating a fault with the connected
device.
Units for <variable label> has
changed. Unit must be sent
before editing, or invalid
data will be sent.
The engineering units for this variable have been edited. Send engineering units
to the device before editing this variable.
Unsent data to online
device. SEND or LOSE data
There is unsent data for a previously connected device which must be sent or
thrown away before connecting to another device.
Upgrade 275 Field
Communicator software to
access XMTR function.
Continue with old
description?
The communicator does not contain the most recent Rosemount 3051N Device
Descriptors (DDs). Select
YES
to communicate using the existing DDs. Select
NO
to abort communication.
Use up/down arrows to
change contrast. Press
DONE when done.
Gives direction to change the contrast of the HC display.
Value out of range
The user-entered value is either not within the range for the given type and size
of variable or not within the min/max specified by the device.
<message> occurred
reading/writing <variable
label>
Either a read/write command indicates too few data bytes received, transmitter
fault, invalid response code, invalid response command, invalid reply data field,
or failed pre- or post-read method; or a response code of any class other than
SUCCESS
is returned reading a particular variable.
<variable label> has an
unknown value. Unit must
be sent before editing, or
invalid data will be sent.
A variable related to this variable has been edited. Send related variable to the
device before editing this variable.
Message