Specifications and Reference Data
October 2016
Reference Manual
00809-0100-4808, Rev DA
44
Specifications and Reference Data
A.3.5 Span and zero, zero elevation, and suppression
Zero and span values can be set anywhere within the range limits stated in
Table A-2
and
Table A-3
, providing sensor limits are not
exceeded.
Span must be greater than or equal to the minimum span stated in
Table A-2
and
Table A-3
.
A.3.6 Humidity limits
0-100% relative humidity
A.3.7 Volumetric displacement
Less than 0.005 in
3
(0,08 cm
3
)
A.3.8 Turn-on time
Two seconds maximum
A.3.9 Response time
Dead time (T
d
)
Maximum dead time before analog output reacts to step change in
input = 0.1 seconds
Update rate
20 times per second minimum
Minimum time constant (T
c
)
Adjustable damping
Time constant on analog output is incrementally adjustable from
the minimum values stated above to 25.6 seconds nominal.
Figure A-1. Typical Smart Transmitter Response Time
Table A-2. Rosemount 3051ND and 3051NG Range and Sensor Limits
Range
Minimum span
Range and sensor limits
Rosemount 3051ND/NG
Upper (URL)
Lower (LRL)
Rosemount 3051ND
Rosemount 3051NG
0
0.1 inH
2
0 (25 Pa)
3.0 inH
2
0 (750 Pa)
-3.0 inH
2
0 (-750 Pa)
N/A
1
0.5 inH
2
0 (0,12 kPa)
25 inH
2
0 (6,22 kPa)
-25 inH
2
0 (-6,22 kPa)
N/A
2
2.5 inH
2
0 (0,62 kPa)
250 inH
2
0 (62,2 kPa)
-250 inH
2
0 (-62,2 kPa)
-250 inH
2
0 (-62,2 kPa)
3
10 inH
2
0 (2,48 kPa)
1000 inH
2
0 (248 kPa)
-1000 inH
2
0 (-248 kPa)
0.5 psia (3,5 kPa abs)
4
3 psi (20,7 kPa)
300 psi (2 070 kPa)
-300 psi (-2 070 kPa)
0.5 psia (3,5 kPa abs)
5
20 psi (138 kPa)
2000 psi (13 800 kPa)
-2000 psi (-13 800 kPa)
0.5 psia (3,5 kPa abs)
Table A-3. Rosemount 3051NA Range and Sensor Limits
Range
Minimum span
Range and sensor limits
Upper (URL)
Lower (LRL)
1
0.3 psia (2,07 kPa abs)
30 psia (206,8 kPa abs)
0 psia (0 kPa abs)
2
1.5 psia (10,34 kPa abs)
150 psia (1 034,2 kPa abs)
0 psia (0 kPa abs)
3
8 psia (55,16 kPa abs)
800 psia (5 515,8 kPa abs)
0 psia (0 kPa abs)
4
40 psia (275,8 kPa abs)
4000 psia (27 580 kPa abs)
0 psia (0 kPa abs)
At 70 °F, with minimum damping setting
Range code
Minimum time constant (Tc) including dead
time (Td)
0
1.0 seconds
1
0.5 seconds
2 - 5
0.2 seconds
At 70 °F, with minimum damping setting
Range code
Minimum time constant (Tc) including dead
time (Td)
20mA
9.89mA
4mA
Time
Transmitter 4–20 mA Output vs. Time
63.2% of Total
Step Change
T
d
= Dead Time
T
c
= Time Constant
Response Time =
T
d
+T
c
Pressure Released
T
c
T
d