96
Using AMS Device Manager to configure the meter
Section 4: Configuration
Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual
June 2017
3-9000-765 Rev H
Figure 4-15 Configuration changes dialog
c.
Click the
Service Tools|Variables
tab. The Variables page displays tabs for the
device’s Flow Data, Path Information, Flow Totals, and All Variables).
Figure 4-16 AMS Device Manager - Service Tools
d.
The
Service Tools|Flow Data
page includes charts for flow and sound velocities.
The flow values (flow direction, average flow velocity and average sound
velocity) parameters are displayed for the connected device.
e.
Click
Service Tools|Variables|Path Information
tab to view the device’s chord
performance (%), Gain (dB), SNR (dB), Signal (mV) and Noise (mV).
f.
Click
Service Tools|Variables|Flow Totals
to view the volume totals (Forward and
reverse Uncorrected Volume) parameters for the connected device.