94
Using AMS Device Manager to configure the meter
Section 4: Configuration
Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual
June 2017
3-9000-765 Rev H
13.
Click the
Flow Analysis
tab to select Configure Reverse Flow Detection, if desired. The
default setting is
Disabled
. Click the
Disabled
button to send the feature command to
the meter. Check for a response error. If no error response is received, click the
Enable
button.
a.
Enter the minimum reverse flow velocity above which to accumulate flow in the
reverse direction for this alert. Enter a positive value for the Reverse Flow Zero
Cutoff. Click the
Next
button to write the values to the meter. Check for an error
response. If no error response is received, click the
Next
button. The Detect
Reverse Flow enabled page displays. Click the
Next
button to display Detect
Reverse Flow disabled.
b.
If an error message is returned, click the
Next
button to display the Method
Complete page.
c.
Click the
Set Flow Range Limits
button and enter a positive value for the Flow
Analysis Lower Velocity Range and the Upper Velocity Range Limits. When the
velocity is outside of the limit parameters, an alert is triggered. Click the
Next
button to display the Method Complete page.