90
Using AMS Device Manager to configure the meter
Section 4: Configuration
Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual
June 2017
3-9000-765 Rev H
AMS Device Manager - Manual Setup
Use the
Manual Setup
wizard to configure the meter’s parameters.
and from the AMS Device Manager Configure menu click
Manual Setup
.
Figure 4-8 AMS Device Manager - Configure Manual Setup
Procedure
1.
If installed, remove security wires from the endcap and the Bracket/Cover hex head
bolts that secures the Base Enclosure.
1.
Disable the Write Protect switch in the CPU Module to write any of the following
configuration parameters to your meter.
2.
Click the
Device Variables Mapping
tab. The Primary and Secondary variables are read
only and are configured for Uncorrected Flow Rate. The Third and Fourth variable
configuration choices include Pressure and Temperature.
3.
Click the
Units
tab
(see
AMS Device Manager - Guided Setup
).
4.
Click the
Analog Output 1 (HART)
tab (see
AMS Device Manager - Guided Setup, Step
).
5.
Click the
Analog Output 2
tab. Follow the configuration instructions in the
Manager - Guided Setup, Step 2b
. The read only Secondary variable Content,
Uncorrected Flow Rate, displays. Use the drop-down arrow and select the (flow)
Direction - Forward or Reverse. Enter a Lower and Upper Range limit. Set the Alarm