58
100-214-294 REV. 05
How it works:
• Fill the tank with warm water and a cleaning solution. The tank can be filled with any amount of
diluted solution as long as it reaches the water level line once the items to be cleaned and
accessories are placed into the tank
• Pour your solution medium into one or more beakers or into a solid insert tray
• Place the beakers in a beaker positioning cover or a solid insert tray to fit your unit. Beakers
should not touch the tank’s bottom
The advantages of this method are:
• Removed soil stays in the beaker or tray so you can easily examine, filter or discard it
• You can use one or more solutions at the same time (two completely different cleaning solutions
or one beaker or tray with a cleaning solution and one with a rinse solution)
• Cleaning solution in your tank needs to be changed less often
6.6.3
Non-Cleaning Application
This method can be used for sample preparation, which includes degassing liquids,
mixing, homogenization, dissolving solids, cell lysing and dispersion or particles.
Figure 6.8
Non Cleaning Application
How it works:
• Fill the tank with water and wetting agent. The tray or beakers can be filled with any amount of
solution as long as the water in the tank outside the tray or beakers reaches the water level line
• Place the beakers in a beaker positioning cover, an insert tray or an insert basket to fit your unit,
or place the beakers and flasks onto a Branson support rack. Beakers should not touch the tank’s
bottom
Summary of Contents for Branson CPX1800
Page 4: ...iv 100 214 294 REV 05 ...
Page 6: ...vi 100 214 294 REV 05 ...
Page 8: ...viii 100 214 294 REV 05 ...
Page 21: ...100 214 294 REV 05 11 Chapter 3 Delivery and Handling 3 1 Unpacking Your Unit 12 ...
Page 35: ...100 214 294 REV 05 27 Chapter 5 Installation and Setup 5 1 Installing Your Unit 28 ...
Page 76: ...72 100 214 294 REV 05 ...