100-214-294 REV. 05
47
6.5.2
CPXH Series LCD Description
Table 6.16 LCD Description for CPXH Series.
Item
Function
Sonics/Degas Set Time
Displays the set time for a timed ultrasonic or degas cycle.
• Press Up and Down keys to adjust ultrasonic or degassing cycle time
(adjustable from 1 to 99 minutes). In Constant Sonics Mode, “- - -” is
displayed
Set Temperature
Displays the target temperature.
Temperature units are indicated by the °F (for Fahrenheit) or °C (for
Celsius) right of the Current Temperature icon.
Current Temperature
Displays the current tank temperature as measured by the unit.
Temperature units are indicated right of the icon as either °F (For
Fahrenheit) or °C (for Celsius). Units can be switched using the Fn key.
See Fn key description on
Sonics/Degas Timer
Displays the remaining time of a running timed ultrasonic or degas
cycle.
Press up and down keys to adjust ultrasonic or degassing cycle time
(adjustable from 1 to 99 minutes).
Auto
Indicates the unit is in Auto Mode. In Auto Mode, the following actions
are carried out automatically by the controller:
• Heater is turned on to bring bath to set temperature
• When set temperature is reached, ultrasonics are activated. The unit will
abort the auto cycle and flash the icon if set temperature is not reached
within a 120-minute period
• When ultrasonics timer reaches 0 minutes, the auto cycle is finished
If at any point during an auto cycle the degas key is pressed, a degas
cycle will begin. If ultrasonics has already started, the ultrasonics timer
will restart after the degas period.
Constant Sonics
Indicates the unit is operating in Constant Sonics Mode. In Constant
Sonics Mode, ultrasonics will remain on until the Sonics key is pressed
or the unit is turned off.
Summary of Contents for Branson CPX1800
Page 4: ...iv 100 214 294 REV 05 ...
Page 6: ...vi 100 214 294 REV 05 ...
Page 8: ...viii 100 214 294 REV 05 ...
Page 21: ...100 214 294 REV 05 11 Chapter 3 Delivery and Handling 3 1 Unpacking Your Unit 12 ...
Page 35: ...100 214 294 REV 05 27 Chapter 5 Installation and Setup 5 1 Installing Your Unit 28 ...
Page 76: ...72 100 214 294 REV 05 ...