Rosemount Model 8800C Vortex Flowmeter
A-12
Flanges
316/316L stainless steel
Collars
Hastelloy C-22
©
Process Connections
Mounts between the following flange configurations:
ASME B16.5 (ANSI): Class 150, 300, 600, 900
DIN: PN 10, 16, 25, 40, 64, 100, 160
JIS: 10K, 20K, and 40K
Mounting
Integral (Standard)
Electronics are mounted on meter body
Remote (Optional)
Electronics may be mounted remote from the meter body.
Interconnecting coaxial cable available in nonadjustable 10, 20, and
30 ft (3,0, 6,1, and 9,1 m) lengths. Consult factory for non-standard
lengths up to 75 ft (22,9 m). Remote mounting hardware includes a
polyurethane painted, carbon steel pipe mount bracket with one
carbon steel u-bolt.
Pipe Length Requirements
The vortex meter may be installed with a minimum of ten straight pipe
diameters (D) upstream and five straight pipe diameters (D)
downstream.
Rated accuracy is based on the number of pipe diameters from an
upstream disturbance. An additional 0.5% shift in K-factor may be
introduced between 10 D and 35 D, depending on disturbance. For more
information on installation effects, see Technical Data Sheet
00816-0100-3250.
Surface Finish of Flanges and Collars
Standard: 125 to 250
m
inches
(3.1 to 6.3
m
meters) Ra roughness
Smooth: 63 to 125
m
inches
(1.6 to 3.1
m
meters) Ra roughness
Summary of Contents for 8800A
Page 1: ...Model 8800C and Model 8800A Smart Vortex Flowmeter 00809 0100 4003 English Rev JA ...
Page 2: ......
Page 4: ...Rosemount Model 8800C Vortex Flowmeter ...
Page 8: ...Rosemount Model 8800C Smart Vortex Flowmeter 4 ...
Page 10: ...Rosemount Model 8800C Vortex Flowmeter 1 2 ...
Page 48: ...Rosemount Model 8800C Vortex Flowmeter 2 38 ...
Page 74: ...Rosemount Model 8800C Vortex Flowmeter 3 26 ...
Page 102: ...Rosemount Model 8800C Vortex Flowmeter 4 28 ...
Page 128: ...Rosemount Model 8800C Vortex Flowmeter B 2 ...
Page 129: ...B 3 Approvals ...
Page 130: ...Rosemount Model 8800C Vortex Flowmeter B 4 ...
Page 131: ...B 5 Approvals ...
Page 132: ...Rosemount Model 8800C Vortex Flowmeter B 6 ...
Page 133: ...Appendix C 1 Appendix C HART Communicator Connections and Hardware page C 6 ...
Page 142: ...Rosemount Model 8800C Vortex Flowmeter C 10 ...
Page 159: ......