![Emerson 8800A Product Manual Download Page 108](http://html.mh-extra.com/html/emerson/8800a/8800a_product-manual_103482108.webp)
Rosemount Model 8800C Vortex Flowmeter
A-6
Load Limitations
Maximum loop resistance is determined by the voltage level of the
external power supply, as described by:
NOTE
HART Communication requires a minimum loop resistance of 250
ohms.
Optional LCD Indicator
Displays flow variable, percent of range, current output, and/or
totalized flow
Enclosure Rating
NEMA Type 4X; CSA Type 4X; IP66
PERFORMANCE
SPECIFICATIONS
Accuracy
Includes linearity, hysteresis, and repeatability
Liquids—for Reynolds Numbers over 20000
R
max
=
41.7(V
ps
– 10.8)
V
ps
=
Power Supply Voltage (Volts)
R
max
=
Maximum Loop Resistance (Ohms)
Power Supply (Volts)
Loa
d (
O
hm
s
)
Operating
Region
1250
1000
500
0
10.8
42
42
Digital and Pulse Output
±0.65% of rate
Analog Output
Same as pulse output plus an additional 0.025% of span
Summary of Contents for 8800A
Page 1: ...Model 8800C and Model 8800A Smart Vortex Flowmeter 00809 0100 4003 English Rev JA ...
Page 2: ......
Page 4: ...Rosemount Model 8800C Vortex Flowmeter ...
Page 8: ...Rosemount Model 8800C Smart Vortex Flowmeter 4 ...
Page 10: ...Rosemount Model 8800C Vortex Flowmeter 1 2 ...
Page 48: ...Rosemount Model 8800C Vortex Flowmeter 2 38 ...
Page 74: ...Rosemount Model 8800C Vortex Flowmeter 3 26 ...
Page 102: ...Rosemount Model 8800C Vortex Flowmeter 4 28 ...
Page 128: ...Rosemount Model 8800C Vortex Flowmeter B 2 ...
Page 129: ...B 3 Approvals ...
Page 130: ...Rosemount Model 8800C Vortex Flowmeter B 4 ...
Page 131: ...B 5 Approvals ...
Page 132: ...Rosemount Model 8800C Vortex Flowmeter B 6 ...
Page 133: ...Appendix C 1 Appendix C HART Communicator Connections and Hardware page C 6 ...
Page 142: ...Rosemount Model 8800C Vortex Flowmeter C 10 ...
Page 159: ......