EES-101 Series Ultrasonic Flowmeter
Revision: 3.0.1
Page 27 of 65
Totalizer Multiplier Options
The totalizer multiplier will affect the display mode of
flow cumulant (positive, negative, net) and the output
cumulant mode of RS485 MODBUS. The following
options are available:
0.
x 0.001 (1E-3)
1.
x 0.01
2.
x 0.1
3.
x 1
4.
x 10
5.
x 100
6.
x 1000
7.
x 10000(1E+4)
Factory default factor is x1
ON/OFF POS Totalizer
ON/OFF POS Totalizer. "ON" indicates the
Flowmeter starts to totalize the value. When it is
turned off, the positive totalizer displays in Window
M02 will not change. Factory default is " ON ".
ON/OFF NEG Totalizer
ON/OFF NEG Totalizer. " ON " indicates the
Flowmeter starts to totalize the value. When it is
turned off, the negative totalizer displays in Window
M03 will not change. Factory default is " ON ".
Totalizer Reset
Totalizer reset; all parameters are reset. Press
;
press
or
arrow to select corresponding
options. And then press
to clear ZERO. The
following options are available:
0.
None:No reset;
1.
All:Reset all totalizers;
2.
NET Totalizer Reset;
3.
POS Totalizer Reset;
4.
NEG Totalizer Reset;
5.
Reset:back to the factory default
If the user wants to delete all the already set
parameters and set back to the factory default, select
reset in this window and then the flow meter will reset
to be the factory default automatically.
Menu
3
3
Totalizer Multiplier
0. x0.001(1E-3)
Menu
3
5
POS Totalizer [35
ON
Menu
3
6
NEG Totalizer [36
ON
Menu
3
7
Enter
^
^
Enter
Totalizer Reset? [37
Selection