40
ELSTER Handel GmbH
MCYC Measurement cycle time
Measurements (e.g. temperature), computed values (e.g. conversion factor) and
counter readings are updated on this cycle.
To ensure all functions,
MCYC
must only be set to integer factors of 60 seconds,
e.g. 5, 10, 15, 20, 30 or 60 seconds. In addition
MCYC
must be an integer factor of
OCYC
(see below).
Changing this value is only possible with the WinPADS parameterisation software.
Entering values which do not fulfil these conditions is rejected with the error mes-
sage "6".
In applications subject to official calibration
MCYC
must be less than or equal to
20 seconds.
The factory setting is 20 seconds.
With settings less than 20 seconds the battery service life is reduced.
(
→
B-2)
DAYb Day boundary
The day boundary indicates how many hours after 0:00 a change of day is to be
carried out.
ADJ.T Clock adjustment factor
ADJ.T
is the deviation of the running accuracy of the clock at room temperature in
per mil (
ž
10
-3
). The TC210 uses
ADJ.T
to optimise the running accuracy of the
clock. The adjustment of the clock is carried out in the factory.
Provided no value has been entered for
ADJ.T
, the TC210 displays the message
11 in the system status in the status
Stat
. (
→
3.3.1.1)
Summary of Contents for TC210
Page 2: ...Temperature Volume Corrector TC210 2 ELSTER Handel GmbH ...
Page 8: ...Temperature Volume Corrector TC210 8 ELSTER Handel GmbH ...
Page 51: ...TemperatureVolume Corrector TC210 ELSTER HANDEL GmbH 51 ...
Page 52: ...Temperature Volume Corrector TC210 52 ELSTER Handel GmbH ...
Page 62: ...Temperature Volume Corrector TC210 62 ELSTER Handel GmbH ...
Page 63: ...TemperatureVolume Corrector TC210 ELSTER HANDEL GmbH 63 Appendix ...
Page 66: ...Temperature Volume Corrector TC210 66 ELSTER Handel GmbH A 2 Certificate for Ex Zone 1 ...
Page 67: ...TemperatureVolume Corrector TC210 ELSTER HANDEL GmbH 67 ...
Page 68: ...Temperature Volume Corrector TC210 68 ELSTER Handel GmbH ...