SW 5250A•SW 3500A•SW 1750A
General Description
Operation Manual
1-9
1.6
1.7
Protection and Safety
Overvoltage Shutdown
: Programmable for 15V to 255V peak, 156V range;
30V to 510V peak, 312V range.
Undervoltage Shutdown
: Automatic, not programmable.
Programmable Current Limit Shutdown
: Can be set to 1% of range;
0.5A to 13A for 156V range, 0.5A to 6.5A for 312V range.
Programmable Current Limit with Timed Shutdown
: Can be set to 1% of range;
the timeout can be set from 100 msec to 10 sec.
Programmable Constant Current
: Can be set to 1% of range;
0.5A to 13A for 156V range, 0.5A to 6.5A for 312V range.
For all current accuracies, ±1% of full scale, add ±1.5%/kHz above 500 Hz.
For paralleled amplifiers, add ±1%.
Temperature Coefficient:
<200 ppm/°C.
Overtemperature Shutdown:
Automatic, not programmable.
Agency
Requirements
The SmartWave is designed to meet the following agency requirements:
•
UL 3111
•
CUL 1010
•
EN 61010
•
EN 50081-2
•
EN 50082-1
•
IEC 555-2
•
IEC 801-2, -3 and -4
•
FCC Part 15, Class A