Page 3
2.3 Ultra-Violet Light
The
MEDICA EDI 15/30
system is fitted with an ultra-violet lamp. The
UV lamp is enclosed in a stainless steel chamber ensuring the
operator will not be exposed to UV light.
WARNING!
LIGHT
FROM THE
UV
LAMP
IS
EXTREMELY HARMFUL TO THE EYES
AND SKIN! UV LAMPS SHOULD ONLY
BE USED IN THE REACTION CHAMBER
WITH SUITABLE PROTECTION CAPS
FITTED. PERSONS SHOULD NEVER BE
EXPOSED TO LIGHT FROM THE LAMP.
2.4 Sanitization Chemicals
No sanitization chemical is supplied with the
MEDICA EDI 15/30
products. The relevant recommended chemical can be sourced
through your local supplier or distributor.
2.4.1. Use of
MEDICA EDI 15/30
within United States
During the sanitization cycle EfferSan
™ Multi-purpose
Disinfecting Tablets is used and relevant safety guidance is
included in this document. Please refer to the manufacturer
for safety data sheets.
EfferSan
™ is EPA registered as a sterilant, high level
disinfectant, and sanitizer.
EfferSan
™ is acidic and requires normal neutralization as
specified by your local state and local regulations.
2.4.2. Use of
MEDICA EDI 15/30
within Rest of the World
During the sanitization cycle a CT3 sanitization tablet is used
and relevant safety guidance is included in this document. A
safety data sheet conforming to COSHH regulations is
available upon request.
2.5 Control of Substances Hazardous to Health (COSHH)
Material safety data sheets covering various aspects of the product
are available where applicable upon request. Contact your local
supplier or distributor.
2.6 Environment
The
MEDICA EDI 15/30
should be installed on a flat, level surface, in
a clean, dry environment. The system can also be wall mounted
against a vertical wall capable of supporting the weight. For this we
recommend the use of the wall mounting kit.