8.2 Replacing the LC212 dual cartridge pack
The dual cartridge pack should be replaced in the following
circumstances:
The water purity alarm monitor indicates that water purity
has fallen below limits.
If the system is being re-commissioned or sanitized after
an extended period in which it has not been used.
When indicated by the change reminder.
Immediately before and after use of the emergency
bypass.
Step 1
– Switch System Off
1. SWITCH the
MEDICA EDI 15/30
off at the power
switch at the top left hand side of the system.
2. RELIEVE any residual pressure from the system
by waiting several minutes before proceeding.
3. OPEN the door and IDENTIFY the LC125 0.2
m
capsule filter.
4. Slowly OPEN the bleed nipple on the 0.2µm
capsule filter until water flow ceases.
5. USE an absorbent cloth to soak up the water and
RE-TIGHTEN the bleed nipple.
WARNING!
ENSURE THE SYSTEM IS ISOLATED
BEFORE REMOVING THE
CARTRIDGE PACK.
Step 2 - Remove Cartridge Pack
1. OPEN the front door.
2. PUSH on cartridge top cap.
3. LIFT up and SLIDE out cartridge.
4. UNSCREW
thumbwheel
to
remove
lower
manifold.
5. REMOVE and DISCARD o-rings from lower
manifold.
6. DISCARD used cartridge.
Note: The consumable is non-hazardous. Dispose of as
ordinary waste, observing all local and national
regulations.
Step 3 - Replace Cartridge Pack
1. REMOVE the new LC212 cartridge pack from its
packaging.
2. REMOVE the sealing plugs from the inlet and
outlet ports of both end caps.
3. INSTALL new o-rings supplied to the lower
manifold spigots.
4. ATTACH lower manifold to the bottom cap ports
and secure using thumbwheel.
5.
WET ‘O’ rings and SLIDE new cartridge into
system.
6. POSITION cartridge onto upper spigots, PUSH
into system.
CAUTION!
Ensure neither of the lower manifold
tubes are not restricted by
installation of the pack.
Removing Cartridge Pack
Replacing Cartridge Pack