MEDICA EDI 15/30
Operator Manual
ELGA
Page 10
MEDICA EDI 15/30
Version 1 05/16
Electrical
The systems can be connected to any electrical supply in the
range of 100 - 240V and 50 - 60Hz. The mains lead is supplied
with a molded plug on one end and a molded connector to the
system on the other. The system should be connected to an
earth.
Drain
A semi-rigid flexible connection to a sink or suitable drain
capable of handling at least 2 l/min is required. The drain point
should be within 10 meters (33 ft) and have a gravity fall at a
maximum of 1 meter (3 ft) above the system. Any connections
directly coupled to drain should have an air-break device fitted.
WARNING!
IN COMMON WITH OTHER ELECTRO-
DEIONIZATION SYSTEMS THIS SYSTEM
PRODUCES
SMALL
AMOUNTS
OF
HYDROGEN IN THE DRAIN LINE. IF THIS
IS ALLOWED TO BUILD UP IN A
RESTRICTED
ENVIRONMENT
THIS
COULD BECOME A HAZARD.
Feed Water
The feed water should comply with specifications provided. This
should enter the system
via an 8mm (5/16”) O/D semi rigid
tube, and should be in the temperature range 1 to 35°C.
CAUTION!
Feedwater temperatures outside the range
1 to 35°C will cause damage to the
MEDICA EDI 15/30
system.
For pressurized feeds, the maximum inlet pressure is 2 bar
(30 psi). Higher feedwater pressures must be reduced using a
pressure reducing valve (Part No. LA 512).
Reservoir feeds to the
MEDICA EDI 15/30
system should be
positioned at the same height, or above the system, to provide
a positive flooded inlet pressure.
Electrical Connections