ELGA
MEDICA EDI 15/30
Operator Manual
MEDICA EDI 15/30
Version 1 05/16
Page 33
5. TURN OFF the electrical supply.
6. WAIT several minutes for residual pressure to
dissipate.
7.
APPLY a suitable warning label such as “DO NOT USE /
CONTAINS STERILANT”.
8.
REMOVE the LC212 dual cartridge pack
(See
Section 8.2 for details).
9. INSTALL the LC213 sanitization bypass cartridge
pack
(follow LC212 procedure
– See Section 8.2
for details).
10. TURN ON the electrical supply. The system will
display the sanitization stand-by screen.
11. PRESS the PROCESS button.
12. If not already displayed, PRESS
button until
RO sanitization mode is shown.
13. PRESS
✔
to accept. The screen will prompt for
the introduction of sanitant chemical to the
system.
Step 2 - Insert chemical into sanitization port
1. CUT 1 EfferSan
™ tablet into quarters using tablet
cutter.
2. UNSCREW cap on sanitization port.
3. ADD ¼ of EfferSan
™ tablet.
CAUTION!
Ensure the correct dosage of chemical is
used.
4.
REFIT cap on sanitization port, hand tight.
Step 3 - Start the sanitization process
1. PRESS the PROCESS button. The sanitization
cycle will last for 20 minutes.
2. Once complete, the display will revert back to the
sanitization stand-by screen.
Note: The sanitization reminder date will be reset
automatically.
3. TURN OFF the electrical supply.
Step 4 - Return to normal operation
1. REMOVE
the
LC213
sanitization
by-pass
cartridge pack.
2. RE-INSTALL the LC212 dual cartridge pack.
3. TURN ON the electrical supply.
4. PRESS the PROCESS button to return to normal
operation.
5. CHECK for any leaks.
6.
REMOVE the “DO NOT USE/CONTAINS STERILANT”
label.
7. STORE or DISPOSE of any remaining chemical as
directed on the manufacturer’s label.
Tablet cutter for cutting
EfferSan
™ tablet.
RO Sanitization mode