ELGA
MEDICA EDI 15/30
Operator Manual
MEDICA EDI 15/30
Version 1 05/16
Page 27
7. ENSURE guide has dropped past retainer.
8.
CLOSE V1 and OPEN V2 to isolate the feed to
the analyzer
(see page 14 for details).
9. SWITCH system on at the power switch.
10. PRESS the PROCESS button to flush the LC212.
11. Slowly OPEN the upper bleed nipple on the
0.2
m capsule filter until water, free of air
bubbles, flows out
– use an absorbent cloth to
prevent splashing of internal components.
12. CLOSE the upper bleed nipple on the 0.2
m
capsule filter.
13. After 5 minutes PRESS the PROCESS button to
stop flow.
14.
OPEN V1 and CLOSE V2 to re-direct flow to
analyzer
(see page 14 for details).
15. PRESS the process button to start the water
purification.
16. CHECK system for leaks.
17. CLOSE the front door.
Note:
If the reservoir is below 70% full, the
system will flush to drain for 30 minutes
and then start to fill the reservoir. If the
reservoir is above 70%, the 30 minute
flush will take place the next time that the
level drops below 70%.
LC125 0.2
m Capsule
Filter
Upper bleed
nipple