ELGA
MEDICA EDI 15/30
Operator Manual
MEDICA EDI 15/30
Version 1 05/16
Page 43
Safety Features
Power fail safe
Boost pump protection from particulates
Re-circulation pump protection from particulates
Low operating voltage 24V
Audio visual alarms
Adjustable alarm settings
Emergency by-pass
Special Features
Low noise levels - minimum intrusion
Optional printer kit for record of operating parameters
Optional remote display
Intermittent re-circulation "sleep" mode
Optional point of use filter
Technologies
Purification Methods
Adsorption
Reverse Osmosis
Conditioning
Ion Exchange
Ultra Violet Radiation-short wavelength
Electro-deionization
In-line 0.2
m Filtration
Purified Water Specification
MEDICA EDI 15
MEDICA EDI 30
*Make up rate
15l /hr
30 l/hr
*Daily output(nominal max)
Up to 300 Litres / 24 hour day
Dispense rate from tap
1.3 l/min (nominal)
**Output back pressure to tap (max)
0.1 bar (1 psi)
* Minimum flow rate to analyzer
1.5 l/min
Purity: (delivered to analyzer)
Inorganic-Typical
1 to 18 M
.cm @25
C
Total organic carbon(TOC)
<30ppb
**Bacteria
<1 CFU/ml
pH
Effectively neutral
Particles
0.2µm filtration
*
Standard conditions are 2 bar inlet pressure, 0.5 bar back pressure
at 15 degrees centigrade, fed with potable water and a clean pre-
treatment cartridge. Refer to flow tables outside these conditions.
**
Subject to correct operating and maintenance procedures.