13
Error messages
Error message
Cause
Action
On/Off button flashing
Overtemperature
Leave the device to cool down or make the
surrounding environment cooler.
.
Symbol flashing rapidly
Power deviation
Charge the battery. If the error is still present
after taking this action, please contact elexxion
AG service center
Important note:
If you send the device or accessories to elexxion AG, please go to our website at
easc.elexxion.com (RMA) to register that they have been sent.
Please only send disinfected devices or accessories to elexxion AG. Contaminated devices and accessories cannot be
accepted.
14
Calibration
Power calibration takes place during the annual technical safety inspection. Power calibration is not normally
necessary outside these periods. However, if power calibration is necessary because clear power deviations are
observed, the following procedure should be used:
Please note that class 4 laser radiation is emitted during this procedure, as in normal laser operation.
Please observe the warnings given in Section 2. Warning notices.
Ophir Orion / TH with 30A-SH-V1 measuring head (with valid calibration)
Procedure:
Open casing.
Connect circuit board to claros pico setup controller.
Use terminal software to compare target values with actual values.