![elexxion claros pico User Manual Download Page 20](http://html.mh-extra.com/html/elexxion/claros-pico/claros-pico_user-manual_2388344020.webp)
claros pico user manual V 2.1
page 20 of 34
After this, place the handpiece front part centrally on the rear handpiece part. Then press both parts together in the
handpiece longitudinal direction until the interlock ring can be heard to engage.
The fiber can be removed after use. (To do so, ensure that the soft metal tip has been bent straight again or removed
beforehand.) To remove the handpiece front part, the interlock ring must be pulled to the rear. The handpiece front
part can now be removed.
Carefully pull the fiber out of the handpiece front part and have it be disinfected.
The fibers can be changed during operation with the device switched on. Fibres must not be replaced during laser
emission under any circumstances!
Fibers have a limited service life depending on the power output used and the fibre diameter. Fibres can typically be
used up to 15 times. Disinfected fibers can be disposed of in household waste.