![elexxion claros pico User Manual Download Page 11](http://html.mh-extra.com/html/elexxion/claros-pico/claros-pico_user-manual_2388344011.webp)
claros pico user manual V 2.1
page 11 of 34
Table 2: electromagnetic immunity
Immunity tests
IEC 60601 test level
Compliance level
Electrostatic discharge (ESD)
according to IEC 61000-4-2
+/- 6 kV contact discharge
+/- 8 kV air discharge
+/- 4 kV contact discharge
+/- 8 kV air discharge
Radiated RF fields
according to IEC 61000-4-3
80 MHz-2.5 GHz: 10V/m;
not life-supporting
80 MHz-2.5 GHz: 10V/m
Electrical fast transient interference
(burst) according to IEC 61000-4-4
Not applicable
N/A
Conducted radio frequency according
to IEC 61000-4-6
3V/m
3V/m
Magnetic field at frequency of supply
voltage according to IEC 61000-4-8
3A/m
3A/m
Voltage dips an short interruptions
according to EN 61000-4-11
<5 % U / 10 msec
40 % U / 0.1 sec
70 % U / 0.5 sec
<5 % U / 10 msec
40 % U / 0.1 sec
70 % U / 0.5 sec
Table 3: Recommended safety distances between portable and mobile telecommunication devices and the
claros pico system
The claros pico system is designed for use in an electromagnetic environment in which RF interference is controlled.
The user of the claros pico system can help to prevent electromagnetic interference by complying with the minimum
distance between portable and mobile RF telecommunication devices and the claros pico system, depending on the
power output. [distances given in meters]
Rated output of
transmitter [W]
Safety distance
150 kHz – 80 MHz
Subject to
80 MHz – 800 MHz
Transmitted frequency
800 MHz – 2.5 GHz
0.01
0.01
0.01
0.02
0.1
0.03
0.03
0.06
2
0.14
0.14
0.3
10
0.32
0.32
0.64
100
1
1
2
This results in a safety distance of approximately 14 cm for mobile phones (the transmitted power of which is limited
to around 2 watts) in the D1 and D2 band and 0.3 m in the E band. (assumption: 10V/m compliance level; tested item
not life-supporting)