12.5
Technical data base unit:
Power supply:
5V/DC, < 3.0 A
(with Friwo FW7556M/05 mains adapter supplied)
Fuses:
No accessible fuses
Operating mode:
Intermittent operation / 2 min ON / 1 min OFF
Protection class:
I application part type B / IP 00
Wavelength:
808 nm +/- 10 nm
Max power output (pulse):
5 W
Max. power output (CW):
1.0 W
Pulse frequency:
CW – 20.000 Hz
Pulse duration:
26 µs,50 µs, 150 µs
Aiming beam:
650 nm +/- 5 nm / < 1mW
Operating conditions:
Temp. 15°C – 35°C
Rel. humidity 20 – 85 %
Air pressure: 800 – 1100 hPa
Storage/transport conditions:
Temp. 5°C – 50 °C
Rel. humidity 10 – 85 %
Air pressure: 800 – 1100 hPa
12.6
Accuracy of values displayed:
output:
+/- 20 % from value specified