![Elbur B 521 II Instruction Download Page 21](http://html1.mh-extra.com/html/elbur/b-521-ii/b-521-ii_instruction_2376919021.webp)
ELBUR: PB 521 II, PB 526 II
Elbur sp. z o.o. sp.k.
21 of 36
6.11. Assembly of side rails
(does not apply for PB 521 II)
•
push the plastic slider with the point facing upwards
into the metal guide groove on the end board, while
pressing the button on its side (releasing the button
will lock the slider in the top guide);
•
position of holes in the side rail cappings are not
the same:
- drilled holes offset downwards
–
top capping for
bed rail,
- drilled holes offset upwards
–
bottom capping for
bed rail.
The side rails in the foot section end are assembled in the same way
as in the head section. If the bed is equipped with wooden side rails on both
sides of the nursing bed, assembly them in the same manner as the first.
Warning
Wrong assembly of the side rails does not let to achieve required distance
from the top of the mattress to the top edge of the top side rail, which
can caused patient’s entrapment or falling out of t
he bed!
6.12. Assembly of the lifting pole
•
in both corners of the head section of the mattress
support platform there are sockets for assembly
the lifting pole. It should be installed on the side,
where patient gets in or out of the bed.
•
the proper location of the lifting pole is determined
by the horizontal pin inside the socket,
•
the handle strap is attached to the lifting pole.
Patient can hold it to make it easier for get in/out
of the bed.
7.0. Operating of the nursing bed
Before first use of the nursing bed, please read instruction carefully,
especially potential risks described in subclause 2.3, which can appeared
if safety procedures, instructions of using and operating are not followed.