ELBUR: PB 521 II, PB 526 II 

 

Elbur sp. z o.o. sp.k. 

 

 

13 of 36 

 

Make sure when adjusting the bed that no patient, operator or any other 
people, especially children can be jammed or injured. 

 

The  nursing  bed  must  be  cleaned  and  washed  down  with  disinfectant               
in case of a patient change or in other situation required it. 

 

There  must  be  a  minimum  height  of  22  cm  without  compression 
measured from the top of the mattress to the top edge of the top side 
rail. This is illustrated on the pictured below. 

 

 mattress support platform, 2 

 foot end board, 3 

 head end board, 4 

 mattress, 

G ≥ 22cm

 

 

When  different  mattress  thicknesses  are  in  use  (thicker  than  12  cm),             
operator or person in charge has a duty to decide to undertake special 
safety measures in order to prevent trapping or person completely falling 
out in accordance to EN 60601-2-52. Do not hesitate to ask for an advice 
Elbur customer service.  

2.3.4. Assembly information 

Warnings 

 

Read instruction carefully before commencing assembly.  

 

Verify  the  shipment  contents  with  your  order  before  commencing 
assembly  to  make  sure  that  you  have  got  all  necessary  parts                         
and components.  

 

Remember to use proper tools.  

 

Two people qualified in nursing bed’s operating are required to assembly.

 

 

There  must  be  enough  space  for  bed  height  adjustment.  It  cannot  be 
blocked by any obstacles (bedside lockers or windowsills). 

 

Make sure that cables connected control box with hand switch and other 
drive  units  are  not  damaged,  that  there  is  no  loop  in  the  cables               
and that they are not crushed by moving parts.

 

2.3.5. Service information 

Warnings 

 

Only  specially  authorized  technicians  are  allowed  to  carry  out 
troubleshooting activities and replace individual electric components.  

 

Use the test protocol draft on pages 28 

 31 during maintenance (you 

can  find  it  also  on  website  www.elbur.eu) 

  the  subsequent  technical 

safety check has to be documented.  It has to be done at least once  a 
year and every re-use at the new place. 

Summary of Contents for B 521 II

Page 1: ...INSTRUCTION FOR USE ELBUR Nursing Beds PB 521 II PB 526 II ELBUR PB 521 II ELBUR PB 526 II...

Page 2: ......

Page 3: ...ty of our products In order to meet customer expectations we provide wide range of bed height adjustment We also care about attractive design as well as matching furniture systems and accessories whic...

Page 4: ...es 11 2 3 1 General information 11 2 3 2 Electrical components information 12 2 3 3 Information for care personnel 12 2 3 4 Assembly information 13 2 3 5 Service information 13 3 0 Scope of delivery 1...

Page 5: ...6 12 Assembly of the lifting pole 21 7 0 Operating of the nursing bed 21 7 1 Electric mattress support adjustment 22 7 2 Side rails operating 23 7 3 Setting the springiness of mattress support platfor...

Page 6: ...rective 93 42 EEC concerning medical devices EN 60601 1 Medical electrical equipment Part 1 General requirements for basic safety and essential performance EN 60601 2 52 Medical electrical equipment P...

Page 7: ...20 Conditions operating storage Temperature o C 10 40 Humidity 30 75 Temperature o C 5 45 Humidity 30 75 Mattress Dimensions HxWxL 10x90x200 cm 12x90x200 cm Volume density of foamed material 35 50 kg...

Page 8: ...barcode max patient weight safe working load manufacturer information and warning symbols 1 3 2 Information labeling Max Patient Weight 140 kg for PB 521 II this value is 165 kg Safe Working Load 175...

Page 9: ...pplication part type B according to EN 60601 1 electrical appliance class II double insulated ingress protection of electrical parts from splashwater do not use the medical device outdoors product mus...

Page 10: ...any time It is necessary to contact physician in charge to make sure that there are not any limitations of use nursing bed in daily healthcare Due to the design ELBUR nursing beds are only approved f...

Page 11: ...in proper and faultless condition Only use technically perfect bed Contact manufacturer or distributor if the exchange or repair of faulty parts have to be made If damage or malfunction is suspected...

Page 12: ...cable is damaged that there is no loop in the cables and that they are not crushed by moving parts or components of other devices It is essential to leave the nursing bed unused for at least 18 minut...

Page 13: ...stomer service 2 3 4 Assembly information Warnings Read instruction carefully before commencing assembly Verify the shipment contents with your order before commencing assembly to make sure that you h...

Page 14: ...contact your distributor or Elbur customer service if you miss any component or get something in bad condition Scissor lift with drive unit Different models are available to order without castors on k...

Page 15: ...done by manufacturer 4 2 Back rest foot rest drive units Nursing beds described in this instruction manual have one drive unit used to adjust back rest section and one drive unit used to adjust upper...

Page 16: ...es and spare parts Warnings Use only original Elbur accessories and spare parts modification of nursing beds is strictly forbidden without authorization of our company It is necessary to perform speci...

Page 17: ...d boards out of the box bring all accessories side rails drive units and all other components out of the box remove all unnecessary package s items 6 3 Assembly of mattress support platform on the sci...

Page 18: ...pport platform scissor lift lugs then connect it using quick release bolts 6 4 Location of drive units and control box for mattress support platform Warning Take special care in assigning the drive un...

Page 19: ...n drive unit is fixed under mattress support platform with the motor housing towards the head end board label on the bed align holes in holders and the drive insert the quick release bolt and secure b...

Page 20: ...t functions ensure that the cables and wires cannot be damaged or overstretched The nursing bed should be plugged in to the separate socket during the tests There must be enough space for bed height a...

Page 21: ...t Warning Wrong assembly of the side rails does not let to achieve required distance from the top of the mattress to the top edge of the top side rail which can caused patient s entrapment or falling...

Page 22: ...adjustment range of height or angle of mattress support platform is possible only within a predetermined range The electrical adjustments can be made using the hand switch which functions are describ...

Page 23: ...rails during lowering them to avoid rapid decrease The ends of side rails shall be always at the same level of mattress support platform head section and foot section Lowering height of mattress suppo...

Page 24: ...different position follow safety rules indicated below mattress support platform has to be in flat and horizontal position side rails has to be latched and in upper position unplug the mains power cab...

Page 25: ...olt out of the hole in the mounting holder then first person can carefully lower the back rest to the flat position To convert the bed back to its normal operating state the quick release bolt has to...

Page 26: ...c components 4 Drive units Housings should be wiped only with a slightly moisturized cloth 9 0 Service Warnings Only specially authorized technicians are allowed to carry out troubleshooting activitie...

Page 27: ...t inspection Repeat inspection Inspection following repairs servicing Manufacturer s final inspection Electrical appliance class I II Manufacturer Elbur sp z o o sp k Bed type PB _____________________...

Page 28: ...ease bolts mounting drive units fitted properly YES NO Is lifting pole distorted YES NO Is triangle grab handle still good enough to use YES NO 1 2 Visual inspection of the electrical components Is th...

Page 29: ...re control and blocking functions of the hand switch working properly YES NO Are drive units working properly abnormal noises smooth operation YES NO 3 2 Performance check of the mechanical components...

Page 30: ...__________________________________________________________ ____________________________________________________________ Comments ____________________________________________________________ __________...

Page 31: ...after 30 minutes break from using adjustment functions Disabler in the hand control activated Deactivate disabler turn key to the left Hand control damaged Notify customer service Control box damaged...

Page 32: ...perfect condition Technical specifications subclause 1 2 contain guidance of operating and storage conditions for ELBUR nursing beds Nursing bed shall be protected from destruction under external infl...

Page 33: ...__________ ____________________________________________________________ ____________________________________________________________ ____________________________________________________________ ______...

Page 34: ...ELBUR PB 521 II PB 526 II Elbur sp z o o sp k 34 of 36...

Page 35: ......

Page 36: ...Elbur sp z o o sp k Dzia osza 34 56 500 Syc w POLAND Tel 48 62 786 97 80 e mail info elbur eu www elbur eu E TF 2E 13 01 02 11 2020...

Reviews: