ALMAG-01
Operating manual Page 57
8 MAITENANCE
8.1 Maintenance of the Device should be conducted by the operating personnel.
8.2 Maintenance procedure is presented in the Table.
Description of job
Periodicity
1.
Visual inspection of the device body and
the mains cable for soundness.
2.
Cleansing, disinfection.
Prior each procedure.
One time a month or when handing the
device over to another user.
9 STORAGE AND TRANSPORTATION
Storage of the Device is in manufacturer package :
- environmental temperature : b40 °C and -50 °C;
- relative humidity at +25 °C: up to 98 %;
- atmospheric pressure: from 84 to 106,7 kPa (630-800 mmHg);
- atmosphere should be free of acid or alkaline vapours as well as of any other
aggressive admixtures.
The Device in the manufacturer's package can be transported by motor, railway or
water (with the exception of sea vessels) air means of transport in accordance with the
transport regulations.
Transport conditions:
- environmental temperature:
b50 °C and -50 °C;
- relative humidity at +25 °C:
up to 100%;
- atmospheric pressure:
from 84 to 106,7 kPa (630-800 mmHg).
The packed Device should be protected from to direct atmospheric precipitation or
mechanical impact.