7. DESCRIPTION OF PRODUCT
7.1.
Usage according to assignation
Compressors are the source of clean, oil-free compressed air determined for connecting to dental appliances
and equipment.
Compressors are made according to the assignation in the following version:
KOMPAKT 5
(5 l. air tank)
KOMPAKT 10
(10 l. air tank)
Dental compressors DK50 Z
and DK50-10 Z
- with base that allows autonomous setting of the
compressor in a space.
Dental compressors DK50 Z/K and DK 50-10 Z/K
- with base and equipped by condensation and filtration
unit (KJF1).
Dental compressors DK50-10 Z / M
- with base and equipped by adsorption air dryer.
Dental compressors DK50 S
and DK50-10 S
- in compact boxes with efficient suppression of noise,
suitable for placing in doctor’s office.
Dental compressors DK50 S/K
and DK50-10 S/K
- in compact boxes and equipped by condensation and
filtration unit (KJF1).
Dental compressors DK50-10 S/M
-
in compact boxes and equipped by adsorption air dryer.
Without additional filtration equipment, the compressed air from compressor is not
suitable for the operation of breathing appliances or similar equipment.
8. DESCRIPTION OF THE FUNCTION:
Compressor
(Fig.1)
Aggregate of compressor (1) sucks in the free air through a line filter (8) and compresses it through a reverse
valve (3) into an air nozzle (2). The apparatus draws the compressed air from the air nozzle, whereby the
pressure drops to closing pressure set on the air-pressure switch (4), at which the compressor is switched
on. The compressor again compresses the air into the air nozzle up to the value of cutoff pressure when the
compressor is switched off. The safety valve (5) prevents the pressure excess in the air nozzle to above
max. of 8 bars. The release valve (7) releases the condensate from the air nozzle. Such compressed and
clean air free from oil traces is in air chamber ready for further usage.
Compressor with dryer
(Fig.2 , Fig.3)
Aggregate of compressor (1) sucks in the free air through a line filter (8) and it is delivered into the air dryer
in compressed form. The air proceeds through the cooler (15) through the dryer chamber (9) with adsorption
apparatus (16), where the humidity is captured through a built-in filter (17) and reverse valve (3) as dried and
clean air into the air nozzle (2). The adsorption apparatus is automatically regenerated in the regeneration
cycle after each compressor switching off by the air-pressure switch. The adsorption chamber is
depressurized through an open solenoid valve (14) at its simultaneous blowing-through with dried air. The
DK50 Z
DK50 S
DK50-10 S
DK50-10 Z
KJF1
MONZUN
M1.10
Summary of Contents for DK50 S
Page 2: ......
Page 3: ...DK 50 06 2008 3 NP DK50_2 OBSAH 4 CONTENTS 18 32 INHALT 47 SOMMAIRE 61 SPIS TRE CI 76 Ed 2 F...
Page 33: ...DK50 06 2008 33 NP DK50_2 1 CE CE 93 42 EEC 2 2 1 2 2 2 3...
Page 34: ...DK50 NP DK50_2 34 06 2008 3 CE 4...
Page 39: ...DK50 06 2008 39 NP DK50_2 3 MONZUN M1 10 4 KJF1...
Page 40: ...DK50 NP DK50_2 40 06 2008 9 5 C 40 C 70 6 10 10 1 5 X Y...
Page 45: ...DK50 06 2008 45 NP DK50_2 15 5 11 1 1 2 3 16 10 1 8 3 2 7 17 1 8 18 11...
Page 46: ...DK50 NP DK50_2 46 06 2008 19...
Page 95: ......