EKO INSTRUMENTS CO., LTD Grating Spectroradiometer MS-713 Instruction Manual Ver5
Pg. 27
Table 7-3. Messages Displayed on the Interval Measurement Dialog Window
Messages
Statuses
Actions
Error (Opening COM Port failed)
Cannot open the setup
communication port.
Check the measurement settings.
Getting the sensor information
Software retrieving the sensor
information. Normal condition
No action necessary
Error (Communication with the
sensor failed)
No responses from the instrument
Check the communication cable
and power supply of the instrument.
Error (Getting the sensor
information failed)
The communication is unstable.
Check the communication cable for
disconnection and cable condition.
Error (Getting the calibration value
failed)
Error (Sensor Name)
A sensor which is different from the
setup is connected.
Check the communication port and
the corresponding instrument.
READY
Completed the measurement
preparation.
No action necessary.
Measuring
Displayed during the measurement
is taken.
No action necessary.
Retrying measurement
The measurement failed, but
retrying to start measurement.
Check communication cable and
power supply of the instrument.
Measurement failed
Displayed when the measurement
is retried for 3 times and
measurement failed.
Measurement OK
Measurement was done
successfully.
No action necessary.
Application Log File (YYYY(Year).LOG)
The YYYY(Year).LOG file is created in the same folder as WSD files saving folder, which was setup in the
measurement setup for Measurement software WSDac. This file records the information such as
application start-up, measurement start, measurement error, file processing error. (Table 7-3)
Table 7-4. Contents Indicated in Application Log File
Message
Condition
APP START
Measurement Software startup
APP CLOSE
Measurement Software exit
INT.MEAS START
Started Interval Measurement
INT.MEAS STOP
Ended Interval Measurement
ERROR MEAS THREAD CREATE
Internal Error (Measurement Thread startup failed)
ERROR WSD LOAD
Existing WSD file uploading for adding data failed
ERROR MS710 MEAS
MS710 Measurement failed.
ERROR MS711 MEAS
MS711 Measurement failed.
ERROR MS712 MEAS
MS712 Measurement failed.
ERROR MS713 MEAS
MS713 Measurement failed.
ERROR WSD SAVE
WSD File Saving failed.
ERROR CSV SAVE
CSV File Saving failed.