EKO INSTRUMENTS CO., LTD Grating Spectroradiometer MS-713 Instruction Manual Ver5 

Pg.  37 

 

8 - 5 .   G r a p h   S e t t i n g  

When specifying the graph axis and color of line on the displayed graph, click the [Graph] button.    Go to 

[View 

 Graph Settings] from the tool bar or click the [Graph] button (Figure 8-11) located at the top then the 

[Graph Settings] window will display as sown on Figure 8-12. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

[Y-Axis Setting]: 

By checking the “Auto” checkbox, the Y-Axis scale is automatically setup.  By 

removing check from the “Auto” checkbox, the maximum value of Y-Axis can be 

specified 

[X-Axis Setting]: 

Specify the X-Axis wavelength range to be displayed on graph. 

[Graph Color, Frame 

Color, and Grid Line]: 

By clicking the colored area, the color selection dialog window will be displayed, 

and the graph and frame color can be specified as desired.    The grid line of the 

graph can be shown or not shown by checking or  un-checking the [Grid Line] 

checkbox. 

[Tiled Data Display Size 

Setting]: 

Setup the size of the one measurement data to be displayed on the Tiled Data 

display window. 

[OK, Apply, Cancel] 

Buttons: 

By clicking the [OK] button, the Graph Setting window will disappear and go back 

to the graph display window. 

Clicking [Apply] button will allow the user to see the changes on the graph with the 

Graph Setting window still opened. 

[Cancel] button can cancel the setting changes on the Graph Setting. 

 

Graph Button 

Figure 8-11. Graph Button

 

Y-Axis Setting 

X-Axis Setting 

Graph Color 

Frame Color 

Grid Line 

Tiled Data Display 

Size Setting 

 

Figure. 8-12. Graph Settings Window

 

Summary of Contents for MS-713

Page 1: ...INSTRUCTION MANUAL MS 713 Grating Spectroradiometer NIR...

Page 2: ...tallation 16 6 2 Data File Format 21 7 Measurement Software WSDac 23 7 1 Starting Measurement Software W SDac 23 7 2 Before Starting Measurement 24 7 3 Start Measurement 26 7 4 Measurement Data Displa...

Page 3: ...EKO INSTRUMENTS CO LTD Grating Spectroradiometer MS 713 Instruction Manual Ver5 Pg 2 10 2 Software 44 10 3 Cables 45 10 4 Dimensions 47 10 5 Accessories List 49...

Page 4: ...el 31 0 70 3050117 Fax 31 0 70 3840607 North America Region www eko usa com info eko usa com EKO INSTRUMENTS USA Inc 111 North Market Street Suite 300 San Jose CA 95113 USA Tel 1 408 977 7751 Fax 1 40...

Page 5: ...713 Instruction Manual Ver5 Pg 4 2 3 About Instruction Manual Copy Rights Reserved by EKO INSTRUMENTS CO LTD Making copies of whole or parts of this document without permission from EKO is prohibited...

Page 6: ...to gale or earthquake and vibration and may invite unexpected accidents Do not use this instrument in different use other than its original purpose Do not disassemble modify or touch inside this instr...

Page 7: ...lding the cover glass dome or any other part of this instrument can lead to damage or dropping instrument Be careful with handling the glass dome 3 2 HIGH LOW TEMPERATURE WARNING 1 Temperature of the...

Page 8: ...h range of 900 nm to 2 500 nm By using MS 713 in combination with MS 711 it is possible to seamlessly measure a broad wavelength range from 300nm to 2 500nm MS 711 has a visible near infrared region b...

Page 9: ...100nm and MS 713 900nm to 2 500nm it is possible to measure a wide wavelength range from 300nm to 2 500nm 3 Operating Software The operating software has different measurement functions to make either...

Page 10: ...s Qty Remarks Sensors 1 Power supply unit 1 Set of Sensor fixing screws 1 Bolts M8 x 230 Washers Nuts 2pcs set Fuse for spare 2 5 2x20 250V 2A Sensor cable 1 Standard Length 10m Ventilation power supp...

Page 11: ...cle size 2 to 3mm Replace when the color changed to red Fuse 1 5 2x20 250V 2A Ventilation 1 Specification of Ventilation Model 9WG1212H101 E Manufacturer SANYO DENKI Size 120 x 120 x 38mm Rated voltag...

Page 12: ...tects the sensor inside from soiling and effects from rain and wind and it enhances the cosine characteristics which incident from the sky hemisphere 5 Sensor Power Supply Cable Connector Connect Sens...

Page 13: ...1 2 TXD RS 422 Red 2 3 RXD RS 422 Orange 3 4 RXD RS 422 Yellow 4 5 Digital GROUND Green 5 6 Supply power 5V Gray Blue 6 7 Supply power ground White Yellow Green 7 DC GND 8 Peltier Control 16V Black 8...

Page 14: ...of reflecting or obstructing surfaces are given next Select a mounting position which it is free from obstructions at 5 above horizon The setup location should be easily accessible for periodic maint...

Page 15: ...ry 4 Adjust Height To take measurements by setting up the MS 713 side by side the sensor surface must be adjusted to the same height level 5 Cover Place the Cover on the instrument and fasten with 3 s...

Page 16: ...e end of power cable to AC power supply connector at the rear of power supply Then connect the other end of power cable to the socket properly Make sure that provided power is AC100 240V 50 60Hz 50VA...

Page 17: ...software on a PC which has administrative and or installation authority 1 Power ON the PC and insert the installation disk to the CD disk drive while the Windows desktop screen is displayed 2 Start t...

Page 18: ...xt step To change the destination folder click Change button and select the appropriate folder NOTE When changing the installing folder User Account Control function becomes active Installing and or d...

Page 19: ...button to go to next step 7 After a while the PC display goes dark and User Account Control dialog window appears Click Yes button to start the software installation 8 As the installation is complete...

Page 20: ...tware 1 Uninstalling from the Programs and Features Access by clicking Control Panel Program and Features Select the program to be uninstalled and right click the popup menu will appear Select Uninsta...

Page 21: ...startup the software to be uninstalled from the CD ROM Proceed according to the messages appear on the window If the software is already installed there appear the options as Modify Repair and Remove...

Page 22: ...format data The spectral radiation data of wavelength interval according to the spectroradiometers specification can be saved in CSV file as well as the spectral radiation data which wavelength interv...

Page 23: ...value that are calculated out from the data in each calculated intervals The PAR PPFD are calculated from the data between 400 to 700nm and Illuminance is calculated by multiplying the standard relat...

Page 24: ...switching multiple units of MS 713 on one computer make sure to do the following Restart the software each time the instrument is switched If the software is not restarted the calibration value saved...

Page 25: ...7 2 then the Measurement Settings window will be displayed as shown on Figure 7 3 COM Port Setup User Comment Integration Range Setup Measurement Time Setup Save Folder for Binary Format Save Checkbo...

Page 26: ...SV format file The file name will be the date and time of the measurement such as YYYYMMDDhh WSD YYYY Year MM Month DD Day and hh Time Wavelength Pitch By selecting the Raw the data is saved with the...

Page 27: ...k the Start button Figure 7 6 located at the top the Interval Measurement window is displayed as sown on Figure 7 7 Table 7 2 Measurement Start Functions Functions Descriptions Current Time The curren...

Page 28: ...The measurement failed but retrying to start measurement Check communication cable and power supply of the instrument Measurement failed Displayed when the measurement is retried for 3 times and measu...

Page 29: ...splayed measurement taken User Comment Displays the User Comment entered in the Measurement Setting Range of Integration Value Displays the spectrum radiation integration value W m2 within the wavelen...

Page 30: ...s Setting Specify the X Axis wavelength range to be displayed on graph Graph Color Frame Color and Grid Line By clicking the colored area the color selection dialog window will be displayed and the gr...

Page 31: ...Go to File Print from the tool bar or click the Print button Figure 7 12 at the top to print the graph Also to see the print preview condition go to File Print Preview from the tool bar Printer can b...

Page 32: ...Analysis Software CSVDisp WSDisp exe is data analysis software and it has main function to view the measured data WSD and save as CSV format files 8 1 Starting Data Analysis Software CSVDisp exe As th...

Page 33: ...ying Data List After the measurement data is uploaded Tiled Data Window appears as shown on Figure 8 3 Only on this window the data can be deleted by selecting data then right click Delete 2 Viewing D...

Page 34: ...window Measured Date Time The date and time of the displayed measurement taken User Comment Displays the User Comment entered in the Measurement Setting Range of Integration Value Displays the spectru...

Page 35: ...left clicking and moving the mouse the values of each intersection position are shown at the right bottom of the graph 4 Displaying Multiple Selected Data When selecting multiple data in a sequential...

Page 36: ...s If more than 10 data are selected the colors are repeated Graph colors can be freely changed by opening the WSDisp ini file as text editor and editing each RGB values of the COLOR TABLE Also by addi...

Page 37: ...on Figure 8 10 CSV Data Waveform Specify the wavelength pitch for when the Saving Measurement Value in CSV Format By selecting the Raw the data is saved with the unique wavelength pitch of the instrum...

Page 38: ...ne By clicking the colored area the color selection dialog window will be displayed and the graph and frame color can be specified as desired The grid line of the graph can be shown or not shown by ch...

Page 39: ...e as CSV Format Save the data of PAR W m2 PPFD umol m2 s ILLUMINANCE lx and integral W m2 per each measured time of the selected measurement data in CSV format Save Measurement Data Save the measureme...

Page 40: ...om the tool bar or click the Print button Fig 8 16 at the top to print the graph Also to see the print preview condition go to File Print Preview from the tool bar Printer can be setup from File Print...

Page 41: ...orizontal position If the sensor is not leveled horizontally measurements will be inaccurate by affecting on cosine azimuth response Check Cable Condition Weekly Check the condition of sensor and comm...

Page 42: ...tainties contributed by the standard lamp calibration set up and spectroradiometer performance properties 9 3 Troubleshooting Check the following items in case of trouble with the instrument If any qu...

Page 43: ...80 Dome material Synthetic Quartz Communication RS422 between sensor and power supply unit Power supply 16VDC 5VDC Maximum Power Consumption 95W Dimension Including cover dome 320 325 H mm Weight Sens...

Page 44: ...x 220 D x 120 H mm Weight 1kg 3 Ventilation Unit Table 10 4 Ventilation Unit Specification Items Details Model Number 9WG1212H101 E Manufacturer SANYO DENKI Dimensions 120 x 120 x 38mm Frame Material...

Page 45: ...ration Environment CPU Memory HHD Capacity Display Resolution Interface Pentium 1GHz or more More than 1GB More than 20GB More than 1024 x 768 CD ROM Drive RS 232C port Software Name Versions Measurem...

Page 46: ...3 Instruction Manual Ver5 Pg 45 10 3 Cables Table 10 6 Cable Specifications Cable Name Details Sensor Cable Cable Type Diameter Outer Diameter Cable end Cable Length MVVS 0 5mm2 x 12pins 10 4mm Y Term...

Page 47: ...Ventilation Unit Power Supply Cable Cable Name Details Ventilation Unit Power Supply Cable Cable Type Diameter Outer Diameter Cable ends Cable Length VCTF 0 75mm2 x 3C 4 8mm AD Adapter side THB387 3P...

Page 48: ...EKO INSTRUMENTS CO LTD Grating Spectroradiometer MS 713 Instruction Manual Ver5 Pg 47 10 4 Dimensions 1 Sensor Part Figure 10 3 Sensor Part Dimensions...

Page 49: ...EKO INSTRUMENTS CO LTD Grating Spectroradiometer MS 713 Instruction Manual Ver5 Pg 48 2 Power Supply Figure 10 4 MS 713 Power Supply...

Page 50: ...Silica Gel 500g Spare Fuse 5 2x20 250V 2A Spare Ventilation Specification of Ventilation Model 9WG1212H101 E Manufacturer SANYO DENKI Size 120 x 120 x 38mm Rated voltage 12V Rated current 0 38A Maxim...

Page 51: ...eko co jp EKO North America 111 North Market Street Suite 300 San Jose CA 95113 USA P 1 408 977 7751 F 1 408 977 7741 info eko usa com www eko usa com EKO Europe Middle East Africa South America Lulof...

Reviews: