Connecting
5.1 Safety information for connection
CuratOR LX491W
Instructions for Use, 02/2018
17
5 Connecting
5.1 Safety information for connection
All safety information and warnings for the device must be observed to ensure danger-free
operation.
CAUTION
Changes to device
Do not make any mechanical or electric changes to the device.
EIZO GmbH is not liable for changes made to the device.
CAUTION
Shielding measures
Follow all shielding measures in accordance with local EMC directives. If these guidelines
are not observed, device malfunction may result.
CAUTION
Grounding
The permissible leakage current is not exceeded during the first fault event in accordance
with EN60601-1. The device is grounded with an additional protective conductor to ensure
the greatest possible electric safety.
CAUTION
Excessive currents, short circuits, and ground faults
In accordance with national standards and regulations, protection against excessive cur-
rents, short circuits, and ground faults must be incorporated into the building installation.
NOTICE
Changes to device settings
Device settings may only be adjusted by service personnel.
NOTICE
Disconnecting from line power
Always set the power switch to "Off" before disconnecting the device from power. Other-
wise the device could be destroyed.