Data Logger Operation
A) Storing the load value currently displayed on the screen
during system operation:
Press
DATA L.
DATA L
will appear on the screen. Press
ENTER
.
The display will flash
OK
and then revert back to the current load
value. The reading has now been stored in the D.L. memory
B) Storing a MAX reading in the D.L. memory:
Along with the current load value, the D.L. can also record the load
value currently stored in the system's MAX memory.
Press
DATA L.
DATA.L
will appear on the screen. Then press
MAX
. The display will flash
OK
and then revert back to the
current load value. The current MAX reading has now been stored
in the D.L. memory. Note: MAX readings will appear as D.L. MAX
under the function (FUNCT) column when output to a PC/printer.
C) Seeing the total number of D.L. entries:
Press
DATA L.
DATA.L
will appear on the screen. Then press
TARE
.
The display will read
DL. N
and then show the total
number of readings currently stored in the D.L. memory.
D) Resetting/clearing the D.L. memory:
Press
DATA L.
DATA.L
will appear on the screen. Then press
RESET
and
RESET
will flash before reverting back to the current
load value. The D.L. memory is now clear and the next load logged
will appear as #1.
E) Downloading the stored info to PC/serial printer:
Note that for a PC, you will need to use a terminal emulator such as
Microsoft's HyperTerminal™.
To output the entire contents of the D.L. memory, press
DATA L.
When the screen reads
DATA.L
press
. All readings should
now be output to the screen/printer.
Ideally, the D.L. data should be output while the handheld indicator is still
connected to the load cell. However, the indicator may be used alone to
output the readings if need be. When turning the indicator on while not