Unauthorized switching of ranges may affect the performance and linearity of the tooling. The tool-
ing should be used with the following magnifications:
Marked Range
Air Amplification
.010” /0.2mm
Low
.005” /0.1mm
Low
.002” / .05mm
Med
.001” / .02mm
Med/High
.0005”/ .01mm
High
To set the range of the module on the bargraph, turn the front panel range switch to required posi-
tion using a small screwdriver inserted thru the front panel.
2-2.3 Mastering
1.) Check the two master sizes and determine the set points on the bargraph scale. With an
air plug, the smaller size is on the bottom. Set the pointers on the side cursor to the 2 set
points. Usually this will be with zero in the middle.
2.) Place the large-sized, or maximum ring gage master over the air plug nozzles.
3.) Adjust the top knob (M) to align the column with the to pointer.
4.) Place the minimum ring gage master over the air plug nozzle.