TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Report on the performance testing of the Serinus 30 ambient air quality moni-
toring system manufactured by Ecotech Pty Ltd measuring CO, Report no.:
936/21221977/D_EN
Page 93 of 253
EP_Ecotech_Serinus30_en.docx
7.5
Assessment
The sensitivity coefficient to the surrounding temperature does not exceed the performance
criteria of max. 0.3 µmol/mol/K. For both systems, the highest value b
st
is used for the pur-
pose of evaluating uncertainty. For system 1 it is 0.040 µmol/mol/K and for system 2 it is
0.046 µmol/mol/K.
Does this comply with the performance criterion? yes
7.6
Detailed presentation of test results
Individual results of the tests are detailed in Table 19.
Table 19:
Individual results of the test of the sensitivity to the surrounding temperature for CO
Device 1
Device 2
Device 1
Device 2
Date
Time
Temp [ºC]
[µmol/mol]
[µmol/mol]
Time
Temp [ºC]
[µmol/mol]
[µmol/mol]
13/05/2013
16:17:41
20
0.0
-0.3
16:32:37
20
65.6
65.8
13/05/2013
16:18:37
20
0.0
-0.3
16:33:33
20
65.6
65.8
13/05/2013
16:19:33
20
0.0
-0.3
16:34:29
20
65.6
65.8
0.0
-0.3
65.6
65.8
14/05/2013
08:05:37
0
0.1
-1.2
08:21:29
0
66.3
64.4
14/05/2013
08:06:33
0
0.1
-1.1
08:22:25
0
66.3
64.4
14/05/2013
08:07:29
0
0.1
-1.1
08:23:21
0
66.3
64.4
0
0.1
-1.1
66.3
64.4
14/05/2013
14:52:41
20
-0.3
-0.5
15:07:37
20
65.4
64.8
14/05/2013
14:53:37
20
-0.3
-0.5
15:08:33
20
65.4
64.8
14/05/2013
14:54:33
20
-0.3
-0.5
15:09:29
20
65.4
64.8
-0.3
-0.5
65.4
64.8
15/05/2013
08:08:49
30
-0.3
-0.6
08:41:29
30
65.3
64.4
15/05/2013
08:09:45
30
-0.3
-0.6
08:42:25
30
65.3
64.4
15/05/2013
08:10:41
30
-0.3
-0.6
08:43:21
30
65.3
64.4
-0.3
-0.6
65.3
64.4
15/05/2013
14:30:49
20
-0.2
-0.9
14:59:45
20
65.8
64.2
15/05/2013
14:31:45
20
-0.2
-0.9
15:00:41
20
65.8
64.2
15/05/2013
14:32:41
20
-0.2
-0.9
15:01:37
20
65.8
64.2
-0.2
-0.9
65.8
64.2
Zero level
Span level
Average (X
1(TS1)
)
Average (X
2(TS2)
)
Average (X
Ts,1
)
Average (X
2(TS1)
) = (X
1(TS2)
)
Average (X
Ts,2
)
Summary of Contents for Serinus 30
Page 120: ...Serinus 30 Kohlenmonoxid Analysator Benutzerhandbuch Version 2 1 www ecotech com...
Page 121: ...Serinus 30 Benutzerhandbuch 2 1 Seite 2 Diese Seite wurde absichtlich frei gelassen...
Page 135: ...Serinus 30 Benutzerhandbuch 2 1 Seite 16 Diese Seite wurde absichtlich frei gelassen...
Page 181: ...Serinus 30 Benutzerhandbuch 2 1 Seite 62 Diese Seite wurde absichtlich frei gelassen...
Page 197: ...Serinus 30 Benutzerhandbuch 2 1 Seite 78 Diese Seite wurde absichtlich frei gelassen...
Page 229: ...Serinus 30 Benutzerhandbuch 2 1 Seite 110 Diese Seite wurde absichtlich frei gelassen...
Page 237: ...Serinus 30 Benutzerhandbuch 2 1 Seite 118 Diese Seite wurde absichtlich frei gelassen...