TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Report on the performance testing of the Serinus 30 ambient air quality moni-
toring system manufactured by Ecotech Pty Ltd measuring CO, Report no.:
936/21221977/D_EN
Page 77 of 253
EP_Ecotech_Serinus30_en.docx
Table 8:
Individual test results for the short-term drift (initial values)
Device 1
Device 2
Device 1
Device 2
Time
[µmol/mol]
[µmol/mol]
Time
[µmol/mol]
[µmol/mol]
06:37:00
-0.1
-0.1
07:08:00
65.2
64.5
06:38:00
-0.1
-0.1
07:09:00
65.2
64.5
06:39:00
-0.1
-0.1
07:10:00
65.2
64.5
06:40:00
-0.1
-0.1
07:11:00
65.2
64.5
06:41:00
-0.1
-0.1
07:12:00
65.2
64.5
06:42:00
-0.1
-0.1
07:13:00
65.2
64.5
06:43:00
-0.1
-0.1
07:14:00
65.2
64.5
06:44:00
-0.1
-0.1
07:15:00
65.2
64.5
06:45:00
-0.1
0.0
07:16:00
65.2
64.5
06:46:00
-0.1
0.0
07:17:00
65.2
64.5
06:47:00
-0.1
0.0
07:18:00
65.2
64.6
06:48:00
-0.1
0.0
07:19:00
65.2
64.6
06:49:00
-0.1
0.0
07:20:00
65.2
64.6
06:50:00
-0.1
0.0
07:21:00
65.2
64.6
06:51:00
-0.1
0.0
07:22:00
65.2
64.6
06:52:00
-0.1
0.0
07:23:00
65.3
64.6
06:53:00
-0.1
0.0
07:24:00
65.3
64.6
06:54:00
-0.1
0.0
07:25:00
65.3
64.7
06:55:00
-0.1
0.0
07:26:00
65.3
64.7
06:56:00
-0.1
0.0
07:27:00
65.3
64.7
Average
-0.1
0.0
Average
65.2
64.6
Start values
Start values
Zero level
Span level
Summary of Contents for Serinus 30
Page 120: ...Serinus 30 Kohlenmonoxid Analysator Benutzerhandbuch Version 2 1 www ecotech com...
Page 121: ...Serinus 30 Benutzerhandbuch 2 1 Seite 2 Diese Seite wurde absichtlich frei gelassen...
Page 135: ...Serinus 30 Benutzerhandbuch 2 1 Seite 16 Diese Seite wurde absichtlich frei gelassen...
Page 181: ...Serinus 30 Benutzerhandbuch 2 1 Seite 62 Diese Seite wurde absichtlich frei gelassen...
Page 197: ...Serinus 30 Benutzerhandbuch 2 1 Seite 78 Diese Seite wurde absichtlich frei gelassen...
Page 229: ...Serinus 30 Benutzerhandbuch 2 1 Seite 110 Diese Seite wurde absichtlich frei gelassen...
Page 237: ...Serinus 30 Benutzerhandbuch 2 1 Seite 118 Diese Seite wurde absichtlich frei gelassen...