TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Seite 44 von 253
Report on the performance testing of the Serinus 30 ambient air quality
monitoring system manufactured by Ecotech Pty Ltd measuring CO, Report
no.: 936/21221977/D_EN
EP_Ecotech_Serinus30_en.docx
6.1
5.2.1 Certification range
The certification range intended for testing shall be determined.
6.2
Equipment
No additional equipment is required to test this criterion .
6.3
Testing
The certification range intended for testing shall be determined.
6.4
Evaluation
VDI Guideline 4202 Sheet 1 and Standard DIN EN 14626 stipulate the following minimum
requirements for the certification ranges of continuous ambient air monitoring systems for
carbon monoxide:
Table 3:
Certification range VDI 4202 Sheet 1 and DIN EN 14626
Measured
component
Lower limit
CR
Upper limit
CR
Limit value
(alert
threshold)
Assessment period
in mg/m³
in mg/m³
in mg/m³
Carbon monoxide
0
100
10
8 h
6.5
Assessment
The measuring system can be assessed in the range of the relevant limit values.
Does this comply with the performance criterion? yes
6.6
Detailed presentation of test results
Not applicable in this instance.
Summary of Contents for Serinus 30
Page 120: ...Serinus 30 Kohlenmonoxid Analysator Benutzerhandbuch Version 2 1 www ecotech com...
Page 121: ...Serinus 30 Benutzerhandbuch 2 1 Seite 2 Diese Seite wurde absichtlich frei gelassen...
Page 135: ...Serinus 30 Benutzerhandbuch 2 1 Seite 16 Diese Seite wurde absichtlich frei gelassen...
Page 181: ...Serinus 30 Benutzerhandbuch 2 1 Seite 62 Diese Seite wurde absichtlich frei gelassen...
Page 197: ...Serinus 30 Benutzerhandbuch 2 1 Seite 78 Diese Seite wurde absichtlich frei gelassen...
Page 229: ...Serinus 30 Benutzerhandbuch 2 1 Seite 110 Diese Seite wurde absichtlich frei gelassen...
Page 237: ...Serinus 30 Benutzerhandbuch 2 1 Seite 118 Diese Seite wurde absichtlich frei gelassen...