TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Seite 14 von 253
Report on the performance testing of the Serinus 30 ambient air quality
monitoring system manufactured by Ecotech Pty Ltd measuring CO, Report
no.: 936/21221977/D_EN
EP_Ecotech_Serinus30_en.docx
Performance criterion
Minimum requirement
Test result
Com-
pliance
Page
5.
Performance characteristics
The manufacturer’s specifica-
tions in the instruction manual
shall be by no means better
than the results of the perfor-
mance tests.
No discrepancies between the in-
strument design and the instruction
5.2
General requirements
Shall meet the requirements of
Table 1 of VDI Guideline 4202
Sheet 1.
The measuring system can be as-
sessed in the range of the relevant
The upper limit of measure-
ment of the measuring systems
shall be greater or equal to the
upper limit of the certification
range.
By default the measuring range is set
to 0 – 100 mg/m³ for CO. Other
200 ppm are possible.
The upper limit of the measuring
range is larger than the respective
upper limit of the certification range.
5.2.3 Negative output signals May not be suppressed (life ze-
ro).
The measuring system also displays
In case of malfunction of the
measuring system or failure in
the mains voltage, uncontrolled
emission of operation and cali-
bration gas shall be avoided.
The instrument parameters
shall be secured by buffering
against loss. When mains volt-
age returns, the instrument
shall automatically reach the
operation mode and start the
measurement.
When mains voltage returns the
measuring system goes back to a
failure-free operational status and au-
tomatically resumes measuring.
The measuring system shall al-
low the control of important op-
erating states by telemetrically
transmitted status signals.
By means of various connectivity op-
tions and the “Serinus Downloader”
software the measuring system can
be monitored and controlled from an
Switch-over between meas-
urement and functional check
and/or calibration shall be pos-
sible telemetrically by computer
control or manual intervention.
In general, all necessary tasks related
to functional checks may be per-
formed directly on-site or monitored
telemetrically using the remote control
Preferably 3 months but at
least 2 weeks.
As determined by the necessary
maintenance tasks the maintenance
Summary of Contents for Serinus 30
Page 120: ...Serinus 30 Kohlenmonoxid Analysator Benutzerhandbuch Version 2 1 www ecotech com...
Page 121: ...Serinus 30 Benutzerhandbuch 2 1 Seite 2 Diese Seite wurde absichtlich frei gelassen...
Page 135: ...Serinus 30 Benutzerhandbuch 2 1 Seite 16 Diese Seite wurde absichtlich frei gelassen...
Page 181: ...Serinus 30 Benutzerhandbuch 2 1 Seite 62 Diese Seite wurde absichtlich frei gelassen...
Page 197: ...Serinus 30 Benutzerhandbuch 2 1 Seite 78 Diese Seite wurde absichtlich frei gelassen...
Page 229: ...Serinus 30 Benutzerhandbuch 2 1 Seite 110 Diese Seite wurde absichtlich frei gelassen...
Page 237: ...Serinus 30 Benutzerhandbuch 2 1 Seite 118 Diese Seite wurde absichtlich frei gelassen...