TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Report on the performance testing of the Serinus 30 ambient air quality moni-
toring system manufactured by Ecotech Pty Ltd measuring CO, Report no.:
936/21221977/D_EN
Page 111 of 253
EP_Ecotech_Serinus30_en.docx
Evaluation
The down times during the field test are listed in Table 31.
Table 31:
Availability of the Serinus 30 analyser
System 1
System 2
Time in the field
h
2244
2244
Down time
h
0
0
Maintenance
h
12
12
Total operating time
h
2232
2232
Total operating time incl.
maintenance
2244
2244
Availability
%
100
100
Maintenance times result from the daily test gas feeding for the purpose of determining the
drift behaviour and the maintenance interval as well as times required to change the internal
Teflon filters in the sample gas line.
7.5
Assessment
The availability is 100 %. This complies with the requirements of Standard DIN EN 14626.
Does this comply with the performance criterion? yes
7.6
Detailed presentation of test results
Not applicable in this instance.
Summary of Contents for Serinus 30
Page 120: ...Serinus 30 Kohlenmonoxid Analysator Benutzerhandbuch Version 2 1 www ecotech com...
Page 121: ...Serinus 30 Benutzerhandbuch 2 1 Seite 2 Diese Seite wurde absichtlich frei gelassen...
Page 135: ...Serinus 30 Benutzerhandbuch 2 1 Seite 16 Diese Seite wurde absichtlich frei gelassen...
Page 181: ...Serinus 30 Benutzerhandbuch 2 1 Seite 62 Diese Seite wurde absichtlich frei gelassen...
Page 197: ...Serinus 30 Benutzerhandbuch 2 1 Seite 78 Diese Seite wurde absichtlich frei gelassen...
Page 229: ...Serinus 30 Benutzerhandbuch 2 1 Seite 110 Diese Seite wurde absichtlich frei gelassen...
Page 237: ...Serinus 30 Benutzerhandbuch 2 1 Seite 118 Diese Seite wurde absichtlich frei gelassen...