Report on the performance testing of the Serinus 30 ambient air quality
monitoring system manufactured by Ecotech Pty Ltd measuring CO, Report
no.: 936/21221977/D_EN
EP_Ecotech_Serinus30_en.docx
7.1
8.4.13 Difference sample/calibration port
The difference between sample and calibration port shall not exceed 1.0 %.
7.2
Test procedure
If the analyser has different ports for feeding sample gas and calibration gas, the difference
in response of the analyser to feeding through the sample or calibration port shall be tested.
The test shall be carried out by feeding the analyser with a test gas with a concentration of
70 % to 80 % of the maximum of the certification range of carbon monoxide through the
sample port. The test shall consist of one independent measurement followed by two individ-
ual measurements. After a period of at least four response times, the test shall be repeated
using the calibration port. The difference shall be calculated according to:
100
×
−
=
∆
t
cal
sam
SC
c
x
x
where
SC
∆
is the difference sample/calibration port;
sam
x
is the average of the measured concentration using the sample port;
cal
x
is the average of the measured concentration using the calibration port;
t
c
is the concentration of the test gas;
SC
∆
shall meet the performance criterion specified above.
7.3
Testing
The test was performed in accordance with the requirements on testing as stipulated in DIN
EN 14626. For test gas feeding the path was controlled by means of a three-way valve be-
tween sample and calibration port.
7.4
Evaluation
The following differences between sample and calibration ports were determined:
System 1 (001): -0.09 %
System 2 (002): -0.17 %
7.5 Assessment
This is in complete compliance with the performance criteria stipulated in Standard DIN EN
14626.
Does this comply with the performance criterion? yes
Summary of Contents for Serinus 30
Page 120: ...Serinus 30 Kohlenmonoxid Analysator Benutzerhandbuch Version 2 1 www ecotech com...
Page 121: ...Serinus 30 Benutzerhandbuch 2 1 Seite 2 Diese Seite wurde absichtlich frei gelassen...
Page 135: ...Serinus 30 Benutzerhandbuch 2 1 Seite 16 Diese Seite wurde absichtlich frei gelassen...
Page 181: ...Serinus 30 Benutzerhandbuch 2 1 Seite 62 Diese Seite wurde absichtlich frei gelassen...
Page 197: ...Serinus 30 Benutzerhandbuch 2 1 Seite 78 Diese Seite wurde absichtlich frei gelassen...
Page 229: ...Serinus 30 Benutzerhandbuch 2 1 Seite 110 Diese Seite wurde absichtlich frei gelassen...
Page 237: ...Serinus 30 Benutzerhandbuch 2 1 Seite 118 Diese Seite wurde absichtlich frei gelassen...