MGR10 A/B/C - Revision : I (for 7.0 software revision)
40
4.8 Measure/Hold function
This key triggers measurement cycles according to the trigger mode ( Single or Continuous – refer
to section 4.3)
In Continuous mode : (this mode is not available when operating on batteries – MGR10A )
Press on
[MES.]
to start permanent measurement cycles , the
« Meas »
display is blinking at
each measurement
Press again on
[MES.]
to store the last measured value on the display. Display indicates
“HOLD”
state .
NOTA :
this state doesn’t interrupt the measurement current , the current continue to flow through
the specimen under test.
Press again on
[MES.]
to resume the memorization of the last measured value and to display
new measurement
In Single mode :
Press on
[MES.]
to start a measurement cycles
This stops the measurement current at the end of the cycle
Press again on
[MES.]
to get the following measurement
WARNING :
when measuring with a 10A current , it is possible to create small sparks during
connection and disconnection of the specimen . To avoid damaging the specimen under test , please
make the connection and disconnection without current flowing .
6
MES.
MEAS.