MGR10 A/B/C - Revision : I (for 7.0 software revision)
69
SENSe:FRESistance:MODE <
SL O W | M E D | F A S T
>
Sets the measurement mode. Slow mode produces approximately two readings per second, gives
the best noise immunity, and all features are available. Medium mode gives a 1.5x to 2x
improvement in speed. Fast mode can produce up to 50 readings per second, but some features are
not available:
Average Current Mode:
Forced to +I mode
Temperature Compensation:
Forced OFF
Settling Algorithm:
Forced OFF
SENSe:FRESistance:MODE?
Returns the measurement mode currently in force. The response is an ASCII string: “SLOW”,
“MED” or “FAST”.
SOURce:CURRent <magnitude>,<current-mode>
Sets the measuring current magnitude and mode according to the following parameters. The
current measurement is aborted.
<
magnitude>
10 - 100
(%)
<
current mode
>
+I, -I, AVE
AVE current mode is not available in FAST mode. Sending this command will generate an
“execution error”. The command is ignored and the microhmmeter will remain in the original
current mode.
The Settling Algorithm will be set to OFF if AVE current mode is selected.
For the MGR10C model, which has fixed current, a magnitude of 10 - 100 must be entered, but
will be ignored.
SOURce:CURRent?
Returns the measuring current magnitude and mode currently in force. The response is of the
following form:
<
magnitude
>,<“
current mode
”>
containing the appropriate parameter values as above. The current measurement is not affected.