7.1. Technical data
Measuring principle laser sensor:
particle counting to the light barrier principle
Measuring range laser sensor:
for particle sizes of: 4, 6, 14 and 21 µm
Particle counting according to:
ISO 4406:99, NAS 1638, SAE AS 4059
Particle sizes:
> 4 µm
(c)
, > 6,0 µm
(c)
, >14 µm
(c)
, > 21 µm
(c)
.
Representation of cc:
ISO 4
– 24, NAS 00 – 12, SAE AS 000 – 12
Measuring accuracy:
± 1 (contamination class)
Calibration master sensor:
ISO MTD in oil (ISO 11171)
Calibration laser sensor:
ISO MTD in oil (ISO 11943)
Max. particle concentration:
24000 particle / ml (sensor)
Sensor flowrate:
100 ml / min
Measuring principle water sensor:
determination of the water saturation based on the
change of capacitance
Measuring range water sensor:
saturation:
0
– 100 %
temperature:
-30...+ 70 °C /
-
22 …+ 158 °F
Suction operating range:
-
0,2…0,2 bar
/ -
2.9…2.9 PSI
Pressure operating range:
1,5…350 bar
/
22…5,000 PSI
Viscosity range:
10…400 mm²/s
/
45…1854 SUS
Operating temperature range:
0…70°C
/
32…158 °F
Ambient temperature range:
0…50°C
/
32…122 °F
Internal temperature range:
0…45°C
/
32…113 °F
Fluid connections:
1 x mini measuring screwed joint M 16x2,
2 x plug-in coupling for hose DN 6
Fluid compatibility:
hydraulic and lubricating fluids based on
mineral oil (also see on separate list of
compatibility (7.7))
Power supply:
11.1 V DC
External power supply unit:
100...240 V AC/ + 15 V DC/ 6 A, 50-60 Hz
Internal rechargeable battery:
15 V (Automatic with main connection)
Protection class:
IP 67 (with closed cover)
Measuring type:
on-line: single, continuous, cyclic
tank samples with internal pump
Cycle interval:
up to 0,5...24 h, adjustable
Display - output:
contamination classes, number of particles,
saturation, temperature, diagrams, graphics