Instruction manual CCS 5, 15/ 10/ 2021
19
2.5.2.3.
Measure-type: cyclic
•
This mode is to accomplish the cyclic measurements.
•
The procedure as described in detail below is performed in any classification type (ISO
4406:99, NAS 1638, SAE AS
4059). (Example: “ISO 4406”).
•
Selection of the
measurement mode of
“cycl.”
For cyclic measurement.
•
Select classification type in
“report standard”.
•
Select
the “flush Time” in
seconds.
•
Select a fluid type if needed.
•
Select the desired cycle
period in hour with [
+
/
-
]
buttons.
•
Select the desired cycle
numbers [2 ~ 100] with [
+
/
-
]
buttons.
•
Confirm the parameter
selection with
.
Cycle period in [h]
The time between one measurement and the next one can be selected between 0,5 hour to
24 hours in half hour steps.
Cycle number
The total number of measurements can be chosen two-steps wise from 2 to 100.
every cyclic measurement will be saved automatically.
•
Start the cyclic
measurement with
[
START
].